NIRPS first light and early science: breaking the 1 m/s RV precision barrier at infrared wavelengths
This paper reports on the performance of the instrument measured in laboratory (results of test plan in Nice and AIV in Paranal) in terms of spectral coverage, dispersion laws and spectral resolutions, and transfer function analysis: instrumental contrast, visibility accuracy, accuracy of the differential phase, of the closure-phase and of the differential visibility. It also provides results of the first tests on sky and the planning of the on-going commissioning.
The respective test setup was designed in collaboration between PTB and the CRIRES+ consortium. The PTB provided optical radiation sources and calibrated detectors for each wavelength range. With this setup, it is possible to measure the absolute efficiency of the gratings both wavelength dependent and polarization state dependent in a wavelength range from 0.9 μm to 6 μm.