Paper
17 September 2007 Laboratory and radiation performance testing results for the e2v model 212 CCD
Author Affiliations +
Abstract
The e2v CCD212 was designed and developed explicitly to support very high accuracy astrometric observations in moderate radiation environments in space. One of the major new innovations in the detector is the use of "charge canals", i.e. regions of raised potential walls rather than notches or channels, in order to mitigate bulk damage effects without the CTI discontinuities associated with spilling over the notch capacity. We report on both pre- and post-radiation performance test results for this detector.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bryan N. Dorland, Roger Foltz, and Augustyn Waczynski "Laboratory and radiation performance testing results for the e2v model 212 CCD", Proc. SPIE 6690, Focal Plane Arrays for Space Telescopes III, 66900K (17 September 2007); https://doi.org/10.1117/12.733974