1.IntroductionAccurate measurement of the radius of curvature (ROC) of a spherical mirror is often critical to the performance of the optical system where the mirror is used. A variety of ways exists to measure the ROC of a spherical mirror.1, 2, 3, 4, 5 The majority of them can be classified into two categories: direct and indirect methods.2 In direct methods, the center of curvature (COC) of the test surface is located, and the distance from it to the surface is measured by either mechanical or optical means.3 In indirect methods, surface sag is measured and the ROC is then calculated from the sag value. Both methods have drawbacks. The direct method requires a large work space for long radii and is susceptible to distance measurement errors, while the indirect method suffers the extreme sensitivity to sag measurement errors. For accurate ROC measurements, a direct method employing a phase-measuring interferometer (PMI) and a distance-measuring interferometer (DMI) is often used.3, 4 The test surface is measured at the cat’s-eye position first and then moved to the confocal position. When the cat’s-eye and the COC are not perfectly aligned with the laser beam focus, the PMI sees no perfect nulls. The longitudinal offsets of the cat’s-eye and the COC from the laser beam focus are defined as null cavity errors,3 which can be calculated from the wavefront errors the PMI measures at each position and compensated in the calculation of the ROC. Two drawbacks are associated with this method. First, the test piece must be moved from the cat’s-eye position to the confocal position. This is rather inconvenient when the test piece is large and heavy. Second, the PMI measurement at the cat’s-eye position is insensitive to the test surface tilt and decenter,4 which makes it difficult to achieve high accuracy in distance measurement by a DMI. In this letter, we present a method to use a laser tracker (see, e.g., Ref. 6) with a PMI to perform ROC measurement on spherical mirrors. Replacement of a DMI with a laser tracker eliminates the aforementioned drawbacks, and high measurement accuracy is achieved. 2.Method DescriptionA laser tracker is essentially a coordinate measurement machine (CMM). It measures polar coordinates instead of Cartesian coordinates. A laser beam is sent out and reflected back by a sphere-mounted retroreflector (SMR). When the SMR is moved, a feed back sensor detects the motion and keeps tracking it. A laser tracker uses a DMI to measure the distance, and two angular encoders to measure the rotation angles. So the laser tracker can be used as a DMI to measure the ROC of a spherical mirror with a PMI, yet its tracking ability makes it much easier to align than a stand-alone DMI. More importantly, compared to the method using a PMI and a DMI, this method does not require the test surface to be moved. Instead, the test surface is fixed at the confocal position with its COC nearly coincident with the focus of the PMI’s laser beam, and the SMR is moved from the PMI beam focus to the test surface while the laser tracker measures the distance it moves. Since the laser tracker measures distances in 3-D space, we have more freedom to place it wherever convenient. In fact, when the laser tracker is put next to the PMI and close to the laser beam focus, the ROC measurement has higher accuracy and the entire testing system is more compact. Figure 1 illustrates the setup of such a measurement system. The measurement procedure is summarized here:
3.ExperimentWe measured the ROC of a concave spherical mirror 0.5 m in diameter and about 2.5 m in ROC. The interferometer, laser tracker, and the spherical mirror were all mounted on the same optical table. Since the laser tracker measures distances with a built-in DMI, it has much higher accuracy than the angle measurement, which is done by angular encoders. To minimize the measurement error of , we put the tracker close to the laser beam focus of the PMI. The tracker’s laser beam was about 90 deg to the interferometer’s optical axis when it tracked the ball at the focus. The position error of the tracker ball due to the angle measurement error gets coupled into the calculation of , but the error was small since the distance from the tracker to the ball was small . With the tracker positioned near the PMI beam focus, the tracker beam was also nearly parallel to the interferometer’s axis when tracking the ball to the mirror. The angle error, in this case, mostly influences the lateral position error of the ball, which has little effect on the accuracy of the measurement. We measured distances from four points on the mirror to the focus and took the average to obtain . All four points were near the edge of the mirror for convenience of mounting the SMR. The SMR can be mounted anywhere as long as it touches the test surface. The measurement results are listed in Table 1. There were about six surface tilt fringes along with the axis across the test mirror aperture during the experiment. The differences within the measurements for the four points can be partly explained by the tilt fringes. Table 1Results of the ROC measurement of a spherical mirror.
We used a Laser Tracker IIPLUS! from Automated Precisions Inc. (see, e.g., Ref. 6). It offers 0.7-arcsec. angle measurement accuracy and 5 ppm absolute accuracy in 3-D spatial measurements. Using these accuracies, along with the estimated errors of the PMI wavefront measurements, we performed a measurement uncertainty analysis. The analysis indicates the measurement uncertainty is less than . The analysis results are summarized in Table 2. Table 2Uncertainty analysis for the ROC measurement listed in Table 1.
For comparison, we measured the same mirror with a Mitutoyo 140 series inside micrometer7 and the same PMI. We obtained an ROC of 2499.915 mm, or less than that from the PMI and laser tracker measurement. We used a combination of six steel rods for this measurement, and the measurement accuracy specified by Mitutoyo is , where is the number of rods and is the maximum distance in millimeters. In our case, and , so the specified accuracy is . The room temperature was 22°C when the measurement was made, while the lengths of the rods are specified at 20°C. The rods are made of tool steel with thermal expansion coefficients . The 2°C difference caused the rods to expand by as much as . After the thermal expansions of the rods were compensated, the measured ROC was . Thus, the PMI and laser tracker measurement is within the uncertainty of the PMI and inside micrometer measurement. 4.ConclusionWe presented a method to use a PMI and a laser tracker to measure the ROC of a spherical mirror. The laser tracker can be placed next to the PMI, which makes the testing system more compact. During the measurements, only the tracker ball is moved from the PMI focus to the test surface and the test piece stays where it is. This is a convenient way to measure a spherical mirror, especially a concave one, with a mid to long ROC. ReferencesD. Malacara, Optical Shop Testing, 460
–464
(1978) Google Scholar
M. C. Gerchman and
G. C. Hunter,
“Differential technique for accurately measuring the radius of curvature of long radius concave optical surfaces,”
Opt. Eng., 19
(6), 843
–848
(1980). 0091-3286 Google Scholar
L. A. Selberg,
“Radius measurement by interferometry,”
Opt. Eng., 31
(9), 1961
–1966
(1992). 0091-3286 Google Scholar
A. Davies and
T. L. Schmitz,
“Defining the measurand in radius of curvature measurements,”
Proc. SPIE, 5190 134
–145
(2003). 0277-786X Google Scholar
Y. Xiang,
“Focus retrocollimated interferometry for long-radius-of-curvature measurement,”
Appl. Opt., 40 6210
–6214
(2001). 0003-6935 Google Scholar
|