19 January 2023 Automatic intraluminal scanning with a steerable endoscopic optical coherence tomography catheter for gastroenterology applications
Author Affiliations +
Abstract

Significance

Endoscopic optical coherence tomography (OCT) enables real-time optical biopsy of human organs. Endoscopic probes require miniaturization of optics, which in turn limits field of view. When larger imaging areas are needed such as in the gastrointestinal tract, the operator must manually scan the probe over the tissue to extend the field of view, often resulting in an imperfect scanning pattern and increased risk of missing lesions. Automatic scanning has the potential to extend the field of view of OCT, allowing the user to focus on image interpretation during real-time observations.

Aim

This work proposes an automatic scanning using a steerable OCT catheter integrated with a robotized interventional flexible endoscope. The aim is to extend the field of view of a low-profile OCT probe while improving scanning accuracy and maintaining a stable endoscope’s position during minimally invasive treatment of colorectal lesions.

Approach

A geometrical model of the steerable OCT catheter was developed for estimating the volume of the accessible workspace. Experimental validation was done using electromagnetic tracking of the catheter’s positions. An exemplary scanning path was then selected within the available workspace to evaluate motion performance with the robotized steerable OCT catheter. Automatic scanning is compared to a teleoperated one and a manual scanning with a nonrobotized flexible endoscope. Spectral arc length, scanning area, spacing between scan trajectories, and time are metrics used to quantify performance.

Results

The available scanning workspace was experimentally estimated to be 255 cm3. The automatic scanning mode provided the highest accuracy and smoothness of motion with spectral arc length of −3.18, covered area of 10.11 cm2, 1.54 mm spacing between 15 sweep trajectories, maximum translation of 27.99 mm, and time to finish of 3.11s.

Conclusions

Automatic modality improved the accuracy of scanning within a large workspace. The robotic capability provided better control to the user to define spacing resolution of scanning patterns.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Oscar Caravaca-Mora, Philippe Zanne, Guiqiu Liao, Natalia Zulina, Lucile Heroin, Lucile Zorn, Michel De Mathelin, Benoit Rosa, Florent P. Nageotte, and Michalina J. Gora "Automatic intraluminal scanning with a steerable endoscopic optical coherence tomography catheter for gastroenterology applications," Journal of Optical Microsystems 3(1), 011005 (19 January 2023). https://doi.org/10.1117/1.JOM.3.1.011005
Received: 1 July 2022; Accepted: 3 January 2023; Published: 19 January 2023
Advertisement
Advertisement
KEYWORDS
Optical coherence tomography

Endoscopy

Endoscopes

Robotics

Optical coherence

Optical microsystems

Teleoperation

Back to Top