1.IntroductionBiological tissues contain a complex mixture of components, including blood, water, fat, and collagen. The quantification of these components has relevance in physiology and diagnosis of disease. For example, melanoma diagnosis can benefit from accurate melanin quantification,1 whereas peripheral arterial disease diagnosis can be improved through precise blood volume and oxygen saturation measurements.2 Traditional methods such as blood sampling or tissue biopsy are invasive and provide information from a small biological sample at a single point in time. Non-invasive probe-based optical spectroscopy can quantify tissue blood content, oxygen saturation, water, lipid, and melanin content but lacks spatial information.3–10 Imaging approaches including planar multi-spectral reflectance imaging,11,12 spatial frequency domain imaging,13–17 and diffuse optical spectroscopic imaging18 suffer from either limited spatial or spectral resolution. Spectral imaging integrates conventional imaging and spectroscopy to attain both high-resolution spatial and spectral information. This approach has enabled morphological and biological tissue analysis for applications including cancer detection, ophthalmology, and microscopy.3,19–29 Multispectral imaging utilizes only a few wavelengths (typically < 10),23,30 whereas hyperspectral imaging20,24–29,31–34 provides full spectroscopic sampling (typically > 100 wavelengths). High-resolution hyperspectral imaging thus generates very large datasets (e.g., 1,000,000 spatial pixels × 100 wavelengths), typically requiring dimensionality reduction and machine learning for analysis.20,24,35 When spectral imaging is based on collecting diffusely reflected light from an object, the spectrum measured at each image pixel depends on the object’s local absorbing and scattering properties. In biological tissue, oxyhemoglobin, deoxyhemoglobin, melanin, water, and other tissue components have optical properties that are tabulated in the literature.36,37 Given a measurement of an object’s diffuse reflectance and knowledge of the absorption and reduced scattering spectra of these pure components, an estimate of the concentration of each component within the measured tissue can be made. This can be done using a variety of fitting algorithms including least squares regression.38–40 However, iterative fitting approaches can often be computationally intensive.39–42 To reduce computation time, machine-learning approaches have been developed for predicting blood volume, oxygen saturation, epidermal thickness, and melanin from hyperspectral images of human skin.41,43 These methods rely on training data generated through analytical or numerical simulations4,42–44 or on hyperspectral image data acquired from in vivo studies. In each of these previous studies, the wavelength-dependent scattering intensity was either assumed or not predicted despite the variation of scattering among tissues and individuals. However, predicting scattering rather than assuming it is crucial as it significantly enhances the accuracy of component predictions by capturing its variability. This approach acknowledges the interplay between scattering and absorption, leading to more reliable and precise estimations of all tissue components. By accurately predicting scattering, the overall model becomes more robust, reflecting the true optical properties of the tissue. Unlike the previously mentioned studies, by incorporating a scattering intensity prediction, this work aims to provide a more comprehensive analysis of tissue properties and extend the analysis beyond commonly studied tissues, such as skin, to include less commonly studied tissues, such as gingiva. Gingiva, or gum tissue, is particularly significant due to its role in oral health. Gingivitis, the inflammation of the gums, is a common condition that can lead to more severe periodontal diseases if left untreated.45 Detecting gingivitis early is crucial as it can prevent the progression to periodontitis, which is associated with tooth loss and other systemic health issues such as cardiovascular disease and diabetes.46,47 Therefore, the ability of our device and model to detect gingivitis can have substantial clinical implications, improving patient outcomes through early intervention and management of oral health. Here, two separate artificial neural networks (ANNs) were trained on simulated data to predict the blood volume, oxygen saturation, melanin content, and scattering intensity of human skin and gingival tissue from simulated diffuse reflectance spectra. The trained model was then applied to real diffuse reflectance spectra obtained from in vivo hyperspectral imaging of tissue. Occlusion was used to induce changes in tissue oxygenation by temporarily removing the arterial supply, abrasion was used to induce inflammation and produce changes in blood content and oxygenation, and subjects with visibly different levels of melanin were imaged to analyze the effects of pigmentation. The method presented here combines high spatial and spectral resolution hyperspectral imaging with neural network analysis to predict in vivo blood volume, oxygen saturation, melanin content, and scattering intensity for both skin and gingival tissues. The increasing availability of robust hyperspectral imaging hardware paired with rapid neural network-based analysis may enable new applications in non-invasive tissue assessment in the clinical setting. 2.Materials and Methods2.1.Hyperspectral Imaging SetupFigure 1 shows the hyperspectral imaging setup used in this study. The camera is a TruTag Technologies Hinalea 4250 () with a 45 to 50 mm focal length lens (f/5.6) providing a 15 deg field of view. This camera uses a Fabry–Pérot interferometer to collect images at 299 wavelengths from 400 to 1000 nm with a full width at half maximum (FWHM) bandwidth of 4 nm. The light source was a Schott DCR III halogen lamp connected to a 110 mm diameter ring illuminator with a mounted linear polarizing film. An orthogonal linear polarizer (Heliopan, P/N 703030, Gräfelfing, Germany) mounted on the camera lens prevents the acquisition of specular reflection from the sample. The extraoral setup includes a bite piece at a working distance of 150 mm, which serves to stabilize the jaw during image acquisition. On the camera side of the bite piece is a set of six calibrated diffuse reflectance standards, mounted at the same working distance as the tissue or sample being imaged. 2.2.Preprocessing and CalibrationThe camera acquires raw images at 450 different Fabry–Pérot mirror spacings. Each spacing permits light at multiple distinct wavelengths to reach the sensor. The TruScope NRT v1.9.1 software provided by the camera manufacturer is used to process the raw dataset into images at each of 299 wavelength bands in the range 400 to 1000 nm, resulting in a data cube (Fig. 2). Illumination non-uniformity was corrected by first capturing a data cube for a uniform 50% diffuse reflectance standard covering the entire field of view. () is the maximum pixel value measured at wavelength . Raw images of unknown objects, , were then corrected for illumination non-uniformity according to Eq. (1), giving a corrected data cube, Diffuse reflectance standards (FOSS) with manufacturer-specified reflectance values of 0.02, 0.10, 0.20, 0.40, and 0.80 were imaged to determine the wavelength-dependent reflectance of the grayscale tiles mounted on the camera side of the bite piece (Fig. 2). The sum of the pixel values within regions were calculated for each of the reflectance standards and each of the grayscale tiles in the field of view. The diffuse reflectance values of the grayscale tiles were then determined based on the values obtained from the reflectance standards. A second calibration relationship was then developed to convert corrected object pixel values to diffuse reflectance at each wavelength using the bite-piece grayscale tiles that are captured in each tissue image. A data cube was acquired for the grayscale tiles and corrected for illumination non-uniformity according to Eq. (1). At each wavelength, a linear least squares fit was applied to the six data points of measured pixel value versus known diffuse reflectance, providing a pair of calibration parameters (slope and intercept) for each wavelength, and . The diffuse reflectance of subsequent objects, , was determined at each wavelength from the corrected pixel value image according to Eq. (2) 2.3.Training and Testing an Artificial Neural Network for Spectral AnalysisEach spatial pixel in a calibrated hyperspectral data cube contains the diffuse reflectance spectrum at the object location being imaged. To estimate the blood volume (), oxygen saturation (), scattering intensity (), and melanin content () at each pixel, an ANN was developed. To train the ANN, simulated diffuse reflectance spectra were generated using two separate Monte Carlo models: one for skin and one for gingiva (Fig. 3).50 First, absorption and reduced scattering coefficients were calculated by selecting a physiologically relevant combination of , , , , and water content ()36,51 and combining these with their respective extinction coefficients, according to Eqs. (3) and (4).52 Rather than the subcutis layer that is present in the skin, dentin and pulp layers were added to simulate the tooth under the gingiva.48,49,53 Both the skin and gingiva models included a constant background absorption in the top two layers.4,40,51 Example spectra are shown in Fig. 4 for the specific choices of , , , , and . The ranges of physiological values were selected to represent a variety of skin pigmentations, skin compositions, and tissue structures.36,37,51,52,54 A tissue’s overall absorption coefficient is given by where represents wavelength, is the concentration, and is the extinction coefficient of absorber . Similarly, the tissue’s reduced scattering coefficient was modeled by where is the scattering intensity and .52These absorption and reduced scattering coefficients were used as inputs to the multi-layer Monte Carlo model, generating a diffuse reflectance spectrum for the specific set of , , , and values. In this work, the tissue refractive index () was held constant at 1.4, epidermal water content was held constant at 50%, and dermal water content was held constant at 70%. For the skin model, the epidermal thickness was held constant at , the dermal thickness was held constant at 2 mm, and the subcutis was semi-infinite.8,37,52,55,56 In the gingiva model, the epithelial thickness was held constant at , lamina propria thickness was held constant at 1.5 mm, dentin was held constant at 2 mm, and the pulp was semi-infinite.48,49 This method was used to generate 30,000 simulated diffuse reflectance spectra, , for each tissue model. Gaussian noise with varying standard deviations was added to the simulated spectra to replicate experimental conditions, resulting in average signal-to-noise ratios ranging from to 16 dB. The 30,000 simulated diffuse reflectance spectra were separated into 19,200 spectra to train the ANN, 4800 spectra to validate the ANN, and 6000 spectra to test the ANN. Due to the output wavelength range from the halogen light source and the optimal collection range of the hyperspectral camera, reflectance spectra were generated in the range of 420 to 830 nm, resulting in a total of 205 wavelengths used as input for the ANN. The diffuse reflectance spectra were scaled at each wavelength across the entire training set using the min–max scaling method provided by the scikit-learn Python library.57 The same scaling method was applied to the , , , and values in the entire training set. Outputs of the ANN were then unscaled to produce the predicted values of the tissue properties. The ANN structure (Fig. 5) consisted of three hidden layers and was tested using the previously mentioned independent set of 6000 simulated diffuse reflectance spectra that were not included in the training step. To determine the size and number of the dense layers in the ANN, the complexity of the input data, output data, and function being fit were all taken into consideration. The size and number of dense layers that were chosen yielded the lowest root-mean-square prediction errors among the networks tested. The rectified linear unit activation (ReLU) was utilized in each hidden layer. The ANN was trained and tested using an AMD Ryzen 7 4700U CPU. 2.4.In Vivo ExperimentsTo test the ability of the ANN to predict , , , and values for tissue in vivo and detect changes in these values, experiments on skin and gingiva were performed on three subjects under an Institutional Review Board approved protocol (U.S.IRB2023CP/04). First, to show the ability to measure changes in oxygen saturation, two fingers were imaged in the same field of view at the same plane as the in-frame grayscale calibration bar (Fig. 1). Blood flow in one finger was occluded for 5 mins using a rubber band before being released, whereas the other finger was unimpeded. Next, the ability to detect small differences in tissue oxygen saturation and melanin content was demonstrated by performing a 10-s finger occlusion in subjects with a range of pigmentation levels. To demonstrate the measurement of physiological changes in gingival tissue, inflammation was induced on one lateral side of the mouth by 1 min of brushing with a stiff-bristle toothbrush, whereas the contralateral side remained unbrushed. Finally, a change in blood volume () was shown with a corresponding decrease in gingivitis over a 3-week timespan from a clinical study investigating oral health. 3.Results3.1.Simulated ExperimentsThe ANN was trained in 4 mins and 38 s on an AMD Ryzen 7 4700U CPU and tested using 6000 diffuse reflectance spectra generated from known , , , and values. Using these reflectance spectra as input, the ANN model predicted , , , and values with low root mean squared error (RMSE) values (Fig. 6). Furthermore, the ANN was compared with Monte Carlo iterative least squares regression using another separate test set of 100 gingiva Monte Carlo simulations with added noise. The ANN and Monte Carlo iterative least squares regression each performed very well with the fitted spectra averaging RMSE of 0.021 and 0.027, respectively. The average time required for the ANN to analyze one spectrum was only 0.07 s on the AMD Ryzen 7 4700U CPU, whereas the Monte Carlo iterative least squares approach took significantly longer, averaging 301.74 s per spectrum on an NVIDIA GeForce RTX 3080 Ti Laptop GPU, making it over 4000 times slower. 3.2.In Vivo ExperimentsFigure 7 shows the results of the first finger occlusion experiment, where a region of interest was selected on both the occluded and non-occluded fingers. The images shown along the top of the plot show the decrease in oxygen saturation of the occluded finger (top finger) over time, whereas the non-occluded finger (bottom finger) remained the same. This is shown in the plot as well where oxygen saturation of the occluded finger gradually decreases until the rubber band is released at the 285-s time point, at which time, the oxygen saturation spikes and begins to return to the initial state. The blood volume also showed elevated levels in the occluded finger immediately after the occlusion began. The melanin content remained relatively unchanged, as expected. Scattering intensity showed small changes before and after occlusion while remaining consistent with the expected in vivo results.52 Figure 8 shows the blood volume, oxygen saturation, and melanin maps predicted using the ANN for a 10-s finger occlusion of three subjects with visibly different levels of pigmentation. The pigmentation differences are further shown in an additional RGB image for each subject. The occluded finger, shown at the top of the images, shows slightly lower oxygen saturation than the non-occluded finger for each subject, as expected. In addition, higher melanin content was measured in the subjects with darker skin pigmentation. By selecting a region of interest on each fingertip, the average blood volume, oxygen saturation, and melanin content were calculated for each subject and can be seen in Table 1. The blood volume, melanin content, and scattering intensity were relatively unchanged between the occluded and non-occluded fingers for each subject. However, the oxygen saturation level in the occluded finger was lower than the non-occluded finger for all subjects. Given that the occlusion only lasted 10 s, the oxygen saturation only decreased by to 5% for each subject. The measured melanin content for subject 1 was the lowest followed by subjects 2 and 3, as expected. Table 1Average tissue component values for regions of interest on occluded and non-occluded fingers.
The effects of abrasion on gingival tissue are apparent in Fig. 9. Teeth and gingiva only on the left side of the images were brushed. The white arrowheads indicate regions where the blood content () is higher on the brushed (left) side of the gingiva than the not brushed (right) side. The increase in blood volume shown in Fig. 9 is an effect of toothbrush abrasion causing inflammation in the gingival tissue. Figure 10 shows the gingival tissue above a tooth afflicted with gingivitis before and after a 3-week regimen of improved oral health. The images show a clear decrease in blood volume, consistent with a decrease in inflammation. 4.DiscussionA hyperspectral imaging system was assembled to capture images at 299 wavelengths between 400 and 1000 nm. The acquisition time for the whole data cube was 25 s, with individual image exposure times of 40 ms, plus a short interval due to tuning of the Fabry–Pérot interferometer. Subject movement during image acquisition was found to be negligible due to the stabilization bar. Illumination uniformity correction and calibration using in-frame reflectance standards ensured accurate calculation of the pixel-by-pixel reflectance spectra. The multi-layer Monte Carlo approach relied on accurate calibration for the detection of changes in the optical properties and tissue components. The unique spectra of the main absorbers in the visible wavelength range (oxyhemoglobin, deoxyhemoglobin, and melanin) were key factors in successful spectral fitting. This was clear in the finger occlusion shown in Fig. 7 where the melanin was relatively unchanged between the occluded and non-occluded fingers, but the oxygen saturation map showed clear differences. By incorporating melanin in the spectral analysis, this system and ANN analysis are suited for imaging skin with a range of pigmentations for tissue component estimation, as demonstrated in Fig. 8. These experiments demonstrated the potential of this system for use in both human skin and gingival tissue in vivo. The finger occlusion tests not only validated the system’s ability to differentiate tissue parameters effectively but also demonstrated the importance of including scattering intensity in the spectral fitting. The scattering intensity showed consistency with expected in vivo measurements, reinforcing the system’s robustness. However, the lack of experimental confirmation that predicting scattering improves model robustness suggests that further validation with in vivo data is necessary. Although this system successfully demonstrated the ability to monitor these parameters in vivo, these results cannot be used to gauge the accuracy of this system due to the lack of an in vivo ground truth. Additional subjects will be necessary to fully assess the clinical relevance and effectiveness of this device. Despite the camera’s ability to collect wavelengths from 400 to 1000 nm, the current ANN used wavelengths from 420 to 830 nm due to the low signal-to-noise ratio obtained at the shorter and longer wavelengths. A different light source that covers the full 400 to 1000 nm range can possibly improve the ANN predictions by incorporating the short wavelengths where blood and melanin absorption are higher while also opening the possibility of including water content in the ANN prediction with longer wavelengths. In addition, other models can be explored, such as random forests, in the future to potentially enhance prediction accuracy, robustness, and speed. The multi-layer Monte Carlo approach for generating ANN training data can also be improved by generating data from multilayered Monte Carlo simulations that include more layers. By doing so, the ANN may be further tailored to tissues with different physiological behaviors. In addition, incorporating training data to model specific conditions and diseases can help capitalize on this method as a fast diagnostic technique. Although neural networks are commonly used for tissue parameter quantification, the specific focus on skin and gingival tissues and the integration of a multi-layer Monte Carlo model with hyperspectral imaging present a novel contribution. This comprehensive approach enhances the accuracy and applicability of the system, providing a robust framework for tissue parameter estimation. Future work will focus on obtaining ground truth values for in vivo data to provide more rigorous validation and developing a more generalized model capable of accurately quantifying tissue parameters for both skin and gingiva. 5.ConclusionThis study reported a unique and versatile method for the prediction of blood volume, oxygen saturation, melanin content, and scattering intensity in both skin and gingival tissue. The hyperspectral camera provides high spatial and spectral resolution. The ANN provides high-speed analysis that can be used in the clinic non-invasively on a variety of human tissues in vivo. Future work aimed at enhancing validation and expanding the model’s applicability to multiple tissue types will further establish the clinical relevance and effectiveness of this imaging system. Code and Data AvailabilityAll relevant code, data, and materials are available from the authors upon reasonable request. Correspondence and requests should be addressed to the corresponding author. AcknowledgmentsThe authors thank the Colgate-Palmolive Company for funding this research and Latonya Kilpatrick for her continuous support and review of this work. ReferencesJ. Chin, E. Wang and M. Kibbe,
“Evaluation of hyperspectral technology for assessing the presence and severity of peripheral artery disease,”
J. Vasc. Surg., 54
(6), 1679
–1688 https://doi.org/10.1016/j.jvs.2011.06.022
(2011).
Google Scholar
Q. Wang et al.,
“Identification of melanoma from hyperspectral pathology image using 3D convolutional networks,”
IEEE Trans. Med. Imaging, 40
(1), 218
–227 https://doi.org/10.1109/TMI.2020.3024923 ITMID4 0278-0062
(2020).
Google Scholar
R. Richards-Kortum and E. Sevick-Muraca,
“Quantitative optical spectroscopy for tissue diagnosis,”
Annu. Rev. Phys. Chem., 47
(1), 555
–606 https://doi.org/10.1146/annurev.physchem.47.1.555 ARPLAP 0066-426X
(1996).
Google Scholar
K. Luchi et al.,
“Spectral-based estimation of components concentration in skin tissue layers with independence of shading via optical modeling of skin tissue,”
Artif. Life Rob., 27
(1), 9
–18 https://doi.org/10.1007/s10015-021-00727-6
(2022).
Google Scholar
R. Van Beers et al.,
“Accurate measurement of and dermal skin hydration using a wearable miniaturized spectrometer,”
Proc. SPIE, 11956 1195602 https://doi.org/10.1117/12.2607346 PSISDG 0277-786X
(2022).
Google Scholar
A. Amelink et al.,
“In vivo measurement of the local optical properties of tissue by use of differential path-length spectroscopy,”
Opt. Lett., 29
(10), 1087
–1089 https://doi.org/10.1364/OL.29.001087 OPLEDP 0146-9592
(2004).
Google Scholar
J. Brown et al.,
“Advances in quantitative UV–visible spectroscopy for clinical and pre-clinical application in cancer,”
Curr. Opin. Biotechnol., 20
(1), 119
–131 https://doi.org/10.1016/j.copbio.2009.02.004 CUOBE3 0958-1669
(2009).
Google Scholar
R. Nachabe et al.,
“Estimation of lipid and water concentrations in scattering media with diffuse optical spectroscopy from 900 to 1600 nm,”
J. Biomed. Opt., 15
(3), 037015 https://doi.org/10.1117/1.3454392 JBOPFO 1083-3668
(2010).
Google Scholar
T. Hanioka et al.,
“Haemoglobin concentration and oxygen saturation in dog gingiva with experimentally induced periodontitis,”
Arch. Oral Biol., 34 657
–663 https://doi.org/10.1016/0003-9969(89)90021-6 AOBIAR 0003-9969
(1989).
Google Scholar
T. Hanioka et al.,
“Changes in hemoglobin concentration and oxygen saturation in human gingiva with decreasing inflammation,”
J. Periodontol., 62 366
–369 https://doi.org/10.1902/jop.1991.62.6.366 JCCBDT 0340-076X
(1991).
Google Scholar
M. Bouchard et al.,
“Ultra-fast multispectral optical imaging of cortical oxygenation, blood flow, and intracellular calcium dynamics,”
Opt. Express, 17
(18), 15670
–15678 https://doi.org/10.1364/OE.17.015670 OPEXFF 1094-4087
(2009).
Google Scholar
J. Shapey et al.,
“Intraoperative multispectral and hyperspectral label-free imaging: a systematic review of in vivo clinical studies,”
J. Biophotonics, 12
(9), e201800455 https://doi.org/10.1002/jbio.201800455
(2019).
Google Scholar
D. J. Cuccia et al.,
“Modulated imaging: quantitative analysis and tomography of turbid media in the spatial-frequency domain,”
Opt. Lett., 30
(11), 1354
–1356 https://doi.org/10.1364/OL.30.001354 OPLEDP 0146-9592
(2005).
Google Scholar
D. Cuccia et al.,
“Quantitation and mapping of tissue optical properties using modulated imaging,”
J. Biomed. Opt., 14
(2), 024012 https://doi.org/10.1117/1.3088140 JBOPFO 1083-3668
(2009).
Google Scholar
S. Gioux, A. Mazhar and D. J. Cuccia,
“First-in-human pilot study of a spatial frequency domain oxygenation imaging system,”
J. Biomed. Opt., 16
(8), 086015 https://doi.org/10.1117/1.3614566 JBOPFO 1083-3668
(2011).
Google Scholar
S. Gioux, A. Mazhar and D. Cuccia,
“Spatial frequency domain imaging in 2019: principles, applications, and perspectives,”
J. Biomed. Opt., 24
(7), 071613 https://doi.org/10.1117/1.JBO.24.7.071613 JBOPFO 1083-3668
(2019).
Google Scholar
B. E. Urban and H. M. Subhash,
“Multimodal hyperspectral fluorescence and spatial frequency domain imaging for tissue health diagnostics of the oral cavity,”
Biomed. Opt. Express, 12
(11), 6954
–6968 https://doi.org/10.1364/BOE.439663 BOEICL 2156-7085
(2021).
Google Scholar
M. B. Applegate and D. Roblyer,
“High-speed spatial frequency domain imaging with temporally modulated light,”
J. Biomed. Opt., 22
(7), 076019 https://doi.org/10.1117/1.JBO.22.7.076019 JBOPFO 1083-3668
(2017).
Google Scholar
A. F. Goetz and W. M. Calvin,
“Imaging spectrometry: spectral resolution and analytical identification of spectral features,”
Proc. SPIE, 834 158
–165 https://doi.org/10.1117/12.942295 PSISDG 0277-786X
(1987).
Google Scholar
G. Lu and B. Fei,
“Medical hyperspectral imaging: a review,”
J. Biomed. Opt., 19
(1), 010901 https://doi.org/10.1117/1.JBO.19.1.010901 JBOPFO 1083-3668
(2014).
Google Scholar
B. Fromy et al.,
“Early decrease of skin blood flow in response to locally applied pressure in diabetic subjects,”
Diabetes, 51
(4), 1214
–1217 https://doi.org/10.2337/diabetes.51.4.1214 DIAEAZ 0012-1797
(2002).
Google Scholar
R. Lugano et al.,
“Tumor angiogenesis: causes, consequences, challenges and opportunities,”
Cell. Mol. Life Sci., 77
(9), 1745
–1770 https://doi.org/10.1007/s00018-019-03351-7 CMLSFI 1420-9071
(2020).
Google Scholar
S. Ortega et al.,
“Use of hyperspectral/multispectral imaging in gastroenterology. Shedding some–different–light into the dark,”
J. Clin. Med., 8
(1), 36 https://doi.org/10.3390/jcm8010036
(2019).
Google Scholar
H. L. Offerhaus et al.,
“Hyperspectral imaging in biomedical applications,”
J. Opt., 21
(1), 010202 https://doi.org/10.1088/2040-8986/aaf2a0
(2019).
Google Scholar
A. Nouvong et al.,
“Evaluation of diabetic foot ulcer healing with hyperspectral imaging of oxyhemoglobin and deoxyhemoglobin,”
Diabetes Care, 32
(11), 2056
–2061 https://doi.org/10.2337/dc08-2246 DICAD2 0149-5992
(2009).
Google Scholar
B. S. Sorg et al.,
“Hyperspectral imaging of hemoglobin saturation in tumor microvasculature and tumor hypoxia development,”
J. Biomed. Opt., 10
(4), 044004 https://doi.org/10.1117/1.2003369 JBOPFO 1083-3668
(2005).
Google Scholar
N. Bedard et al.,
“Multimodal snapshot spectral imaging for oral cancer diagnostics: a pilot study,”
Biomed. Opt. Express, 4
(6), 938
–949 https://doi.org/10.1364/BOE.4.000938 BOEICL 2156-7085
(2013).
Google Scholar
U. Rubins et al.,
“Multimodal device for real-time monitoring of skin oxygen saturation and microcirculation function,”
Biosensors, 9
(3), 97 https://doi.org/10.3390/bios9030097 BISSED 0265-928X
(2019).
Google Scholar
E. Kho et al.,
“Broadband hyperspectral imaging for breast tumor detection using spectral and spatial information,”
Biomed. Opt. Express, 10
(9), 4496
–4515 https://doi.org/10.1364/BOE.10.004496 BOEICL 2156-7085
(2019).
Google Scholar
C. Zakian et al.,
“In vivo quantification of gingival inflammation using spectral imaging,”
J. Biomed. Opt., 13
(5), 054045 https://doi.org/10.1117/1.2982536 JBOPFO 1083-3668
(2008).
Google Scholar
M. P. Grisham, R. M. Johnson and P. V. Zimba,
“Detecting sugarcane yellow leaf virus infection in asymptomatic leaves with hyperspectral remote sensing and associated leaf pigment changes,”
J. Virol. Methods, 167
(2), 140
–145 https://doi.org/10.1016/j.jviromet.2010.03.024 JVMEDH 0166-0934
(2010).
Google Scholar
L. Wan et al.,
“Hyperspectral sensing of plant diseases: principle and methods,”
Agronomy, 12
(6), 1451 https://doi.org/10.3390/agronomy12061451 AGRYAV 0065-4663
(2022).
Google Scholar
D. Saha and A. Manickavasagan,
“Machine learning techniques for analysis of hyperspectral images to determine quality of food products: a review,”
Curr. Res. Food Sci., 4 28
–44 https://doi.org/10.1016/j.crfs.2021.01.002
(2021).
Google Scholar
H. Yu et al., and P. C. Pandey et al.,
“Hyperspectral remote sensing applications in soil: a review,”
Hyperspectral Remote Sensing: Theory and Applications, 269
–291 Elsevier(
(2020). Google Scholar
Y. Garini, I. T. Young and G. McNamara,
“Spectral imaging: principles and applications,”
Cytom. Part A: J. Int. Soc. Anal. Cytol., 69
(8), 735
–747 https://doi.org/10.1002/cyto.a.20311
(2006).
Google Scholar
S. Prahl,
“Optical absorption of hemoglobin,”
http://omlc.ogi.edu/spectra/hemoglobin/index.html
().
Google Scholar
D. J. Segelstein,
“The complex refractive index of water,”
University of Missouri–Kansas City,
(1981).
Google Scholar
A. Nkengne et al.,
“Spectracam®: a new polarized hyperspectral imaging system for repeatable and reproducible in vivo skin quantification of melanin, total hemoglobin, and oxygen saturation,”
Skin Res. Technol., 24
(1), 99
–107 https://doi.org/10.1111/srt.12396
(2018).
Google Scholar
S.-H. Tseng et al.,
“Chromophore concentrations, absorption and scattering properties of human skin in-vivo,”
Opt. Express, 17
(17), 14599
–14617 https://doi.org/10.1364/OE.17.014599 OPEXFF 1094-4087
(2009).
Google Scholar
A. Bjorgan, M. Milanic and L. L. Randeberg,
“Estimation of skin optical parameters for real-time hyperspectral imaging applications,”
J. Biomed. Opt., 19
(6), 066003 https://doi.org/10.1117/1.JBO.19.6.066003 JBOPFO 1083-3668
(2014).
Google Scholar
L. Gevaux et al.,
“Real-time skin chromophore estimation from hyperspectral images using a neural network,”
Skin Res. Technol., 27
(2), 163
–177 https://doi.org/10.1111/srt.12927
(2021).
Google Scholar
I. Fredriksson, M. Larsson and T. Strömberg,
“Machine learning for direct oxygen saturation and hemoglobin concentration assessment using diffuse reflectance spectroscopy,”
J. Biomed. Opt., 25
(11), 112905 https://doi.org/10.1117/1.JBO.25.11.112905 JBOPFO 1083-3668
(2020).
Google Scholar
E. Zherebtsov et al.,
“Hyperspectral imaging of human skin aided by artificial neural networks,”
Biomed. Opt. Express, 10
(7), 3545
–3559 https://doi.org/10.1364/BOE.10.003545 BOEICL 2156-7085
(2019).
Google Scholar
R. Zhang et al.,
“Determination of human skin optical properties from spectrophotometric measurements based on optimization by genetic algorithms,”
J. Biomed. Opt., 10
(2), 024030 https://doi.org/10.1117/1.1891147 JBOPFO 1083-3668
(2005).
Google Scholar
N. Lang, M. Schätzle and H. Löe,
“Gingivitis as a risk factor in periodontal disease,”
J. Clin. Periodontol., 36 3
–8 https://doi.org/10.1111/j.1600-051X.2009.01415.x JCPEDZ 0303-6979
(2009).
Google Scholar
P. Ylöstalo et al.,
“Gingivitis, dental caries and tooth loss: risk factors for cardiovascular diseases or indicators of elevated health risks,”
J. Clin. Periodontol., 33
(2), 92
–101 https://doi.org/10.1111/j.1600-051X.2005.00875.x JCPEDZ 0303-6979
(2006).
Google Scholar
W. Soskolne and A. Klinger,
“The relationship between periodontal diseases and diabetes: an overview,”
Ann. Periodontol., 6
(1), 91
–98 https://doi.org/10.1902/annals.2001.6.1.91
(2001).
Google Scholar
K. L. Vandana and B. Savitha,
“Thickness of gingiva in association with age, gender and dental arch location,”
J. Clin. Periodontol., 32
(7), 828
–830 https://doi.org/10.1111/j.1600-051X.2005.00757.x JCPEDZ 0303-6979
(2005).
Google Scholar
D. Di Stasio et al.,
“Measurement of oral epithelial thickness by optical coherence tomography,”
Diagnostics, 9
(3), 90 https://doi.org/10.3390/diagnostics9030090
(2019).
Google Scholar
M. Bürmen, F. Pernuš and P. Naglič,
“MCDataset: a public reference dataset of Monte Carlo simulated quantities for multilayered and voxelated tissues computed by massively parallel pyxopto python package,”
J. Biomed. Opt., 27
(8), 083012 https://doi.org/10.1117/1.JBO.27.8.083012 JBOPFO 1083-3668
(2022).
Google Scholar
S. Li,
“Human skin characterization and analysis based on hyperspectral reflectance using machine learning,”
Université de Lyon,
(2021).
Google Scholar
S. L. Jacques,
“Optical properties of biological tissues: a review,”
Phys. Med. Biol., 58
(11), R37 https://doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155
(2013).
Google Scholar
A. A. Selifonov et al.,
“Measurement of optical properties of human gums and dentin in the spectral range from 350 to 800 nm,”
Izvestiya Saratov Univ. Phys., 20
(4), 258
–267 https://doi.org/10.18500/1817-3020-2020-20-4-258-267 PHYSGM 1943-2879
(2020).
Google Scholar
S. L. Jacques,
“How tissue optics affect dosimetry of photodynamic therapy,”
J. Biomed. Opt., 15
(5), 051608 https://doi.org/10.1117/1.3494561 JBOPFO 1083-3668
(2010).
Google Scholar
P. Oltulu et al.,
“Measurement of epidermis, dermis, and total skin thicknesses from six different body regions with a new ethical histometric technique,”
Turk. J. Plastic Surg., 26
(2), 56
–61 https://doi.org/10.4103/tjps.TJPS_2_17
(2018).
Google Scholar
D. Lintzeri et al.,
“Epidermal thickness in healthy humans: a systematic review and meta-analysis,”
J. Eur. Acad. Dermatol. Venereol., 36
(8), 1191
–1200 https://doi.org/10.1111/jdv.18123 JEAVEQ 0926-9959
(2022).
Google Scholar
F. Pedregosa et al.,
“Scikit-learn: machine learning in Python,”
J. Mach. Learn. Res., 12 2825
–2830
(2011).
Google Scholar
BiographyThomas T. Livecchi is a PhD student in the Biomedical Engineering Department at Rutgers, The State University of New Jersey, New Brunswick, New Jersey, United States. He received his BS degree in biomedical engineering from the College of New Jersey in 2019. His research focuses on fast, non-invasive optical imaging methods in the ultraviolet to shortwave infrared range of light. Steven L. Jacques has been an investigator in biomedical optics for 40 years. He received his BS degree in biology from MIT in 1975, his MSEE degree in electrical and computer engineer in 1979, and his PhD in biophysics and medical physics in 1984 from the University of California, Berkeley, California, United States. He was a research associate at the Wellman Labs for Photomedicine, 1983–1988; an instructor in dermatology at Harvard Medical School, 1987–1988; assistant and associate professor at the M. D. Anderson Cancer Center, University of Texas, 1988–1996; professor in bioengineering and dermatology at the Oregon Health and Science University, 1996–2017; and currently is an affiliate professor in bioengineering at the University of Washington, Seattle, Washington, United States. Hrebesh M. Subhash is a senior principal scientist and technical lead at the Global Devices and Clinical Method Development Division of Colgate-Palmolive, New York, New York, United States. His research interests and expertise include biomedical optics, optical instrumentation, wearable sensors, and medical image and signal processing, particularly the development of cutting-edge and translational biophotonics technologies that interface and bridge basic engineering research and medical diagnosis and interventions. Mark C. Pierce received his BS degree in physics and his PhD in biomedical optics from the University of Manchester, United Kingdom, in 1997 and 2001. From 2001 to 2005, he completed his postdoctoral training at Wellman Laboratories, Massachusetts General Hospital. From 2005 to 2011, he was a member of the Research Faculty at Rice University. He is currently an associate professor at Rutgers, The State University of New Jersey, New Brunswick, New Jersey, United States. |
Artificial neural networks
Tissues
Scattering
Hyperspectral imaging
Blood
Oxygen
Diffuse reflectance spectroscopy