21 March 2024 Full-aperture extended-depth oblique plane microscopy through dynamic remote focusing
Paolo Pozzi, Vipin Balan, Alessia Candeo, Alessia Brix, Anna Silvia Pistocchi, Cosimo D’Andrea, Gianluca Valentini, Andrea Bassi
Author Affiliations +
Abstract

Significance

The reprojection setup typical of oblique plane microscopy (OPM) limits the effective aperture of the imaging system, and therefore its efficiency and resolution. Large aperture system is only possible through the use of custom specialized optics. A full-aperture OPM made with off the shelf components would both improve the performance of the method and encourage its widespread adoption.

Aim

To prove the feasibility of an OPM without a conventional reprojection setup, retaining the full aperture of the primary objective employed.

Approach

A deformable lens based remote focusing setup synchronized with the rolling shutter of a complementary metal-oxide semiconductor detector is used instead of a traditional reprojection system.

Results

The system was tested on microbeads, prepared slides, and zebrafish embryos. Resolution and pixel throughput were superior to conventional OPM with cropped apertures, and comparable with OPM implementations with custom made optical components.

Conclusions

An easily reproducible approach to OPM imaging is presented, eliminating the conventional reprojection setup and exploiting the full aperture of the employed objective.

1.

Introduction

Light-sheet microscopy is the gold standard method for fluorescence imaging of complex three dimensional samples at high rates.1 Due to the need for multiple perpendicular microscope objectives around the sample, standard light-sheet microscopy has significant geometric constraints for the shape and mounting method of the observed sample. Although this three-dimensional arrangement of sample mounting can sometimes be an advantage,2,3 its use with standard slides and Petri dishes, or in intravital applications in small rodents can be challenging, requiring complex geometries4 for the objective lenses and for sample positioning.

Oblique plane microscopy (OPM)5 is a variant of light-sheet microscopy, in which a single objective lens is used both for light-sheet illumination and fluorescence detection, greatly simplifying imaging of samples optically accessible from a single direction. To achieve this, the excitation light is confined to one edge of the objective aperture so that the light sheet propagates at an angle with respect to the optical axis. Since the illuminated plane is tilted from the optical axis, only a small portion of it would be in focus on a camera in a standard detection path. To have the oblique plane in focus within the whole field of view of the microscope, a re-projection setup is employed in the detection path, with a secondary microscope objective forming a low magnification image of the sample, and a tertiary objective observing such image at an angle.

While this configuration is effective and relatively simple to implement, the aperture mismatch between the secondary and tertiary objectives constitutes a significant drawback, limiting the performance and versatility of OPM. In fact, as the angle between the light sheet and the optical axis decreases, the fraction of fluorescence light coupled in the tertiary objective becomes smaller, reducing both signal intensity and imaging resolution.6 The problem is visually represented in Fig. 1, which shows the effective aperture angles for water immersion objectives on standard OPM for high and low numerical apertures. The scheme assumes the rarely implemented ideal scenario in which the tertiary objective has the same numerical aperture as the primary, which would require the presence of an immersion medium between the secondary and tertiary objective. Nonetheless, it can be observed that a significant portion of the aperture of the primary objective is not employed by the system, up to the extreme scenario of 0.5 NA objectives, in which the two apertures have no overlap, and no fluorescence photons can be detected.

Fig. 1

OPM apertures. Representation of OPM apertures for high and low numerical aperture water immersion objectives. The angle α represents the aperture of the primary objective, β is the light-sheet aperture, and γ is the aperture of an ideal tertiary objective in standard OPM with the same numerical aperture as the primary. The effective aperture of the system is the angle of the overlap between α and γ.

JBO_29_3_036502_f001.png

As a result, only high numerical aperture primary objectives can be used effectively in standard OPM, and the total aperture and resolution of the system remain impaired from the re-projection setup.

This drawback can be mitigated by tilting the direction of propagation of fluorescence light through the use of a carefully positioned discontinuity in refractive index between the secondary and tertiary objectives. This can be achieved either through an immersion chamber with two separate liquids7 or through the use of a specialized axially asymmetric tertiary objective with null working distance.8 However, these modifications can be complex and expensive to achieve, and still require the use of high numerical aperture primary objectives in order to keep the angle between the light-sheet and the optical axis suitably large. OPM with low NA objectives has been proven possible through the use of a diffractive grating between the secondary and tertiary objectives,9 which however introduces constraints on the maximum usable NA, and requires realignment when used with different colors in order to maintain full resolution.

This article presents remote focusing OPM (RF-OPM), an alternative approach to the oblique imaging problem. RF-OPM images the sample through a single objective of arbitrary numerical aperture, presenting no aperture mismatches in the detection path. Instead of using the standard OPM re-projection method, an electrically tunable lens (ETL)10 is introduced in the back focal plane of the optical system, and a complementary metal-oxide semiconductor (CMOS) camera is positioned in the image plane. In this configuration, at any given time, only a narrow section of the oblique plane is in focus on the camera. The position of such narrow section can be dynamically shifted along the oblique plane by changing the focal power of the ETL. Through the linear modulation of the focal power of the lens in sync with the rolling shutter exposure of the camera, images of the oblique plane can be acquired entirely in focus. While limiting the rolling shutter exposure to a small sub-region of the detector significantly reduces the photon budget, the approach is often employed in conventional light-sheet imaging, either to increase axial confinement in confocal light-sheet setups,11 or to enable wide fields of view with high light-sheet numerical aperture.12 These methods are generally considered a standard procedure when imaging very large optically cleared samples13 and are preferred to slit-scanning confocal due to their better optical sectioning and lower exposure to excitation light in the regions of the sample not in focus. Moreover, the use of rolling shutters as confocal slits is also widespread in low photon budget applications such as high speed multifocal imaging14 and superresolution microscopy,15 proving the light loss is not an insurmountable drawback for its imaging use.

2.

Method

2.1.

Working Principle

While RF-OPM can in principle be implemented using any form of incoherent remote focusing, including deformable lenses,16 deformable mirrors,17 Alvarez lenses,18 or optical elements mounted on fast accelerating translation stages,19 this article reports an ETL based design. While the use of a simple quadratic phase device, such as an ETL, is known to be imperfect for remote focusing, especially in high numerical aperture systems,19,20 the use of ETLs is widespread in remote focusing applications in microscopy2123 due to their cost effectiveness and compact form factor. For the same reason, the ETL was employed in the system reported in this article as, while imperfect, it constitutes a relatively inexpensive and readily available device that has performance compatible with the experimental design.

A simplified representation of the RF-OPM optical setup is reported in Fig. 2(b). The setup consists of a conventional epifluorescence microscope, with a 4-f telescope conjugating the objective’s back focal plane with the aperture of an ETL. The conventional dichroic and filter set is then positioned between the ETL and the tube lens.

Fig. 2

Method. (a) Scheme of the imaging geometry in the axial direction for different focal powers of the ETL. (b) Simplified schematic of the optical setup. FP, objective focal plane; OL, objective lens; PP, pupil plane of the system; L1 and L2, 4f telescope; DM1, DM2, and DM3, dichroic mirrors; M1 and M2, image plane mirrors; ETL, tunable lens; CL - cylindrical lens; L3, tube lens; and CAM, camera detector plane.

JBO_29_3_036502_f002.png

The laser light forming the light-sheet is focused by a cylindrical lens in the center of the ETL. Between the two lenses of the 4-f telescope, a pair of dichroic mirrors splits the excitation and detection paths so that a manually adjustable mirror in the image plane of the system can be tilted to move the laser light at the edge of the objective aperture, achieving illumination along an oblique plane.

In this configuration, the light-sheet remains positioned along a fixed oblique plane, but its focus moves along the plane linearly with the focal power of the ETL, as shown in Fig. 2, panel A. Neglecting the chromatic aberrations of the ETL, the camera always remains focused on the horizontal plane intersecting the light-sheet focus. As a result, at any given time, only a thin strip of the camera detects details in focus, with the lateral position of the strip shifting linearly with the ETL focal power. Using a rolling shutter sensor with a short exposure, it is possible to linearly modulate the focal power of the ETL so that the image is always in focus on the exposed pixels. The final output of the detector will therefore be an image of the light-sheet plane [indicated in red in Fig. 2(a)], fully in focus. Three-dimensional images can be acquired by either moving the sample through the light-sheet with a translation stage or, in principle, by conjugating the ETL plane with a galvanometric mirror, as is generally done in OPM.

It should be noted that the camera output obtained in this configuration consists of the projection of the oblique plane in the horizontal direction, and as such the magnification of the final image will be different for the x and y axes of the camera. An appropriate affine transformation should be applied to obtain an unwarped image.

2.2.

Field of View

As discussed in the previous section, the magnification of the system is different along the axes parallel and perpendicular to the propagation direction of the light-sheet. When the light-sheet is correctly aligned at the edge of the pupil aperture, the angle of the beam from the optical axis is

Eq. (1)

α=arcsin(NAobjn)0.5arcsin(NAlsn),
where NAobj and NAls are the numerical apertures of the objective and of the light-sheet beam respectively, and n is the refractive index of the sample. For simplicity in the design of the system, this can also be approximated as

Eq. (2)

α=arcsin(DobjRls2fn),
where f is the focal length of the objective lens; Dobj is the diameter of the back aperture of the objective; and Rls is the radius of the light-sheet beam at the objective back focal plane.

Given the magnification M of the system and assuming the use of a square detector of side D, the dimensions of the field of view of the acquired images is

Eq. (3)

FOV,FOV=DM,DMsinα,
where FOV is the size in the direction perpendicular to the propagation of the light-sheet, and FOV is the one along the propagation direction of the light-sheet. As a consequence, the total axial distance covered by the field of view is

Eq. (4)

FOVaxial=DMtanα.

It is important to notice that the final pixel size along the propagation direction of the light-sheet increases with the tangent of the angle, approaching infinity as the light-sheet becomes vertical. In principle, an extremely wide field of view can be achieved in the axial direction using a very low numerical aperture objective or by moving the light-sheet beam toward the center of the objective’s aperture but very low pixel sampling would be obtained in this situation.

A second limit to the achievable axial size of the field of view is given by the effective dioptric power range of the ETL and its ability to linearly modulate it at a frequency compatible with the desired frame rate. To ensure the most effective use of the ETL focal power range, the telescope conjugating it to the objective back focal plane should precisely match the two apertures. In this configuration, the required achievable focal length (both converging and diverging) of the ETL is

Eq. (5)

ftl=2Mtel2fobj2FOVaxial,
where fobj is the focal length of the objective lens and Mtel is the magnification of the telescope between the objective and the ETL.

2.3.

Resolution and Frame Rate

The estimation of the resolution of a standard OPM can be complex due to the dependence of the effective numerical aperture of the system on the angle between the secondary and tertiary objectives. A complete and exhaustive analysis can be found in literature.24 In principle, RF-OPM images can achieve the nominal resolution of the objective in the oblique plane, and the resolution in the direction perpendicular to the oblique plane is given by the thickness of the light-sheet at its focus. Moreover, at least for low numerical aperture systems, the axial resolution remains constant over the entire field of view since the focus of the lightsheet shifts in sync with the focal plane of the camera. Optimal resolution is however only achieved in a short axial range with high numerical aperture objectives, due to the appearance of spherical aberration when using them to image outside their designed focal plane. This issue could potentially be solved in the future, adopting a remote focusing method more complex than an ETL,19 which would not be affected by spherical aberration.

A limit to both the resolution and frame rate performance is imposed by optical aberrations introduced by the ETL. When mounted horizontally, a conventional ETL introduces aberrations from the desired spherical phase with an RMS amplitude of less than 200 nm, which can generally be neglected in light-sheet microscopy, since typical samples often introduce more severe aberrations.25,26 However, when the ETL is operated at increasing frequencies, secondary modes other than pure defocus begin to be excited, which generally introduce spherical-like aberrations. For readily available commercial ETLs, such secondary modes exhibit a resonant frequency at approximately 400 Hz, therefore precluding applications at extremely high frame rates. At more conventional light-sheet frame rates of tens of Hz, these effects can be considered negligible.

In addition to the excitation of secondary modes in the ETL, a second limitation to the maximum frame rate achievable in RF-OPM is given by the capability of the ETL to linearly modulate focal power over time. If a detector capable of alternating the direction of the rolling shutter at each frame is employed, the ETL can be synchronized with the detector through a triangular waveform. Triangular waveforms are relatively trivial to generate with ETLs, and nonlinear behavior is limited to short intervals between frames in which the direction of the scan is reversed. However, most detectors generally utilized in light-sheet microscopy are not capable of alternating rolling shutter direction at each frame. In this case, the ETL must generate a sawtooth waveform, which presents a critical point at the discontinuity between frames. Since the ETL cannot instantly switch from a large positive focal power to a large negative one or vice versa, a significant interval between frames is necessary to allow the ETL to reset its position. Hence, as the microscope frame rate increases, the fraction of time in which the detector is actually exposed decreases, which could lead to an insufficient signal-to-noise ratio.

2.4.

Experimental Setup

Images were acquired on a custom RF-OPM setup. The setup is designed for a 60X, 1.1 NA water-dipping objective (LUMFL N 60XW, Olympus, Japan), with a back focal plane aperture diameter of 6.6 mm. To prove the versatility of the method, additional experiments were performed with a 20×, 0.5 NA water-dipping objective (UMPlanFL N 20XW, Olympus, Japan).

The back focal plane of the objective is conjugated with the 16 mm aperture of the ETL (EL-16-40-TC-VIS-5D with ECC-1C controller, Optotune, Switzerland) through two lenses of 125 mm and 300 mm focal, respectively (AC254-125-A and AC508-300-A, Thorlabs, United States), forming a 2.4× 4-f telescope. This resulted in perfect matching of the aperture of the 60× objective, whereas the numerical aperture of the 20× objective was cropped down by the ETL diameter to approximately 0.37.

To image green fluorophores, long-pass dichroic mirrors with a cutoff wavelength at around 500 nm are used to split the path between the two lenses of the telescope. Dichroic mirrors with mm-scale thickness are employed (DMLP505L, Thorlabs, United States, 5 mm thick, and Di03-R488-t3-25x36, Semrock, United States, 3 mm thick) in order to minimize aberrations in the light-sheet path introduced by their curvature.

Excitation light is provided by a 40 mW, 473 nm diode laser (λ-beam, RGB laser systems, Germany) coupled to a single-mode fiber. Light from the fiber is collimated to a beam diameter of 8 mm with a reflective collimator (RC08SMA-P01, Thorlabs, United States), focused through a 400 mm focal cylindrical lens (LJ1363RM-A, Thorlabs, United States) in a light-sheet conjugated to infinity by a 150 mm focal length lens (AC254-150-A, Thorlabs, United States), and then reflected through the center of the ETL by a dichroic mirror (DMLP505L, Thorlabs, United States). This configuration led to a final ratio of 0.18 between the aperture diameters of the light-sheet and the objective at the back focal plane of the system.

An additional 0.5× 4-f telescope (AC508-200-A and AC508-100-A, Thorlabs, United States) is present in the fluorescence light optical path between the ETL and the tube-lens of the system. This addition serves two purposes: first, to extend the optical path length, allowing the horizontal mounting of the ETL; and second, to accommodate potential future upgrades, such as the integration of a galvanometric mirror at the system’s back focal plane, which would enable faster dynamic imaging. A fluorescence filter (MF525-39, Thorlabs, United States) is present after the relayed back focal plane, and a 200 mm focal length tube lens (MXA20696, Nikon, Japan) is employed to conjugate the image plane of the system to an sCMOS camera (Orca Flash 4.0 v2, Hamamatsu, Japan), with a final effective magnification of 55.55× when using the 60× objective, and of 16.66× when using the 20× objective. Images were acquired on a 2048 by 1024 pixels subregion of the detector. The active region was cropped vertically to 1024 pixels since the 60× objective showed significant spherical aberration outside this range, whereas the axial field of view achievable with the 20× objective was considered more than adequate for most available samples. Given the geometry of the setup, the final field of view of raw images for the 60× objective span 240  μm on the horizontal plane by 156  μm along the oblique plane, at an angle of 48 deg with the optical axis, for a total axial range of 104  μm. For the 20X objective, the field of view span 720  μm on the horizontal plane by 1020  μm along the oblique plane, at an angle of 20 deg with the optical axis, for a total axial range of 958  μm.

Due to the inability of the detector to image with an alternating direction rolling shutter, a sawtooth function is used for modulating the ETL focal power in time. The maximum usable camera frame rate achievable is 50 Hz, with a duty cycle of exposure of approximately 60%, whereas higher duty cycles can be achieved at lower camera frame rates, up to 90% at 10 Hz. Higher frequencies or better duty cycles could be achieved by vertically cropping the active region of the detector, reducing the axial size of the field of view.

Three-dimensional datasets were acquired by translating the sample horizontally with a servo-controlled actuator (M-405.CG, Physik Instrumente, Germany). A custom software script was used to correct the dataset shearing and stretching through an affine transform.

3.

Results

3.1.

Microbeads Imaging

To evaluate the performance of the system, 3 mm thick samples of 0.17  μm yellow-green fluorescent microbeads (P7220 PS-Speck Point Source Kit, Thermo Fisher Scientific, United States) embedded in agarose gel were imaged with the two objectives. Representative datasets are reported in Fig. 3.

Fig. 3

Microbead imaging. (a) x-z maximum intensity projection over a 100  μm range in y of a microbead image, acquired with the 60× objective. Scale bar is 25  μm. (b) x-z maximum intensity projection (where z is the direction of the optical axis of the objective) over a 25  μm range in y of a microbead image, acquired with the 20× objective. Scale bar is 100  μm. C Estimation of the full width at half maximum of the PSF of the system as a function of axial position with the 60× objective. All bright objects in a 240×420×104  μm3 volume were detected and fitted with a Gaussian function on each axis. Crosses are single bead measurements, lines are average values over 5  μm intervals. The vertical dashed line shows the estimated position of the objective’s focal plane. (d) Estimation of the full width at half maximum of the PSF of the system as a function of axial position with the 20× objective. All bright objects in a 720×540×940  μm3 volume were detected and fitted with a Gaussian function on each axis. Crosses are single bead measurements, lines are average values over 35  μm intervals. The vertical dashed line shows the estimated position of the objective’s focal plane. (e) Maximum intensity projection of the average image of 50 microbeads at each axial position from the dataset in panel (c). Top row is z-y projections, bottom row is x-y projections. Scale bar is 2  μm. (f) Maximum intensity projection of the average image of 50 microbeads for each axial position from the dataset in panel (d). Top row is z-y projections and bottom row is x-y projections. Scale bar is 5  μm.

JBO_29_3_036502_f003.png

A first, important observation is that ETL defocusing with high aperture objectives introduces significant spherical aberration outside of a relatively short axial range of operation. This can be clearly observed in Fig. 3(a) and 3(b), which show how the axial resolution of the system is conserved throughout the axial field of view with the 0.5 NA 20× objective lens but is only optimal within a range of approximately 15  μm for the 1.1 NA 60× objective. Within the 15  μm optimal range of the 60X objective, the average full width at half maximum of the psf of the system was 1.01±0.05  μm axially and 0.60±0.03  μm laterally. The resolving power remains acceptable outside the optimal axial range, whereas the axial resolution quickly approaches 3  μm. Due to the lower numerical aperture of the 20× objective, its resolution was more stable, remaining close to its average value of 1.71±0.04  μm over the entire axial range. The Axial resolution gets progressively slightly worse than the average value of 10.5±0.5  μm measured in the 100  μm closer to the coverslip, possibly due to imperfect index matching of the agarose gel with water. With both objectives the resolving power is slightly worse than the nominal diffraction limit of the objective, and a slight star shape can be observed in the PSF. Both these effects are, most likely, due to slight astigmatism introduced in the detection path by the three dichroic mirrors utilized in the system. To avoid the introduction of vignetting in the system at high defocusing power of the ETL, large dichroics (2-inch diameter) were employed, which introduced non-negligible wavefront distortion. Moreover, the employed objectives are not optimized for use with a coverslip, which is necessary for imaging beads in an inverted microscope. A more optimized layout of the system using a coverslip-optimized objective and smaller higher-quality dichroics should solve this issue.

3.2.

Mouse Kidney Imaging

The main advantage of OPM systems over traditional light-sheet microscopy is its ability to image samples conventionally mounted on a microscopy slide. To show this capability, a commercially available and fairly widespread sample, a 16  μm cryostat section of mouse kidney with immunofluorescence staining (Fluocells prepared slide #3, Invitrogen, United States), was imaged with the setup. The glomeruli and convoluted tubules, stained with Alexa 488, were visible with the laser and filter set present in the setup. Due to the thin nature of the sample, only the 60× objective was utilized. Figure 4 shows the acquisition geometry and a representative image of the sample. Raw planes were acquired at 50 Hz, with a 0.4  μm spacing between planes, on a 2048 by 256 pixels sub-region of the detector. The dataset shows good optical sectioning, and resolution performance comparable to those measured in the microbeads test. Image decorrelation analysis27 reported a lateral resolution of 0.72  μm, compatible to the resolution measured with microbeads. The images produced are also comparable to those reported in literature or on commercial microscopes documentation for the same sample. The main drawback of the current setup is the single-channel acquisition, which limits the amount of information that can be retrieved. Future upgrades with multi-edge dichroics could easily enable the use of multiple excitation wavelengths. The widely available dual rolling shutter feature of sCMOS cameras, together with the use of an image splitter, could also allow the simultaneous acquisition of two channels, doubling the acquisition pixel rate of the system.

Fig. 4

Prepared mouse kidney slide imaging. (a) Geometry of the acquisition, not to scale. The green plane represents the objective’s focal plane, the red plane represents the raw field of view of the system, and the blue plane represents the axial field of view of the system. The orange arrow shows the direction of the sample translation. (b) Representative images of the sample. Image border colors report their geometry referring to the scheme in panel (a). Images in the horizontal and vertical plane are obtained through affine transform of the raw data. Scale bar for the horizontal and the vertical image is 50  μm, scale bars in raw image are 20  μm, and intensity scale bar is in arbitrary units.

JBO_29_3_036502_f004.png

3.3.

Zebrafish Imaging

Tg(kdrl:eGFP)s843 zebrafish embryos, expressing a green fluorescent protein in the vascular structure28 were imaged at 3 to 5 days post fertilization (dpf). The pigmentation was suppressed through 1-phenyl-2-thiourea (PTU) treatment.29 To immobilize the larvae, 0.016% tricaine anesthetic solution was used. The inverted nature of the microscope allowed convenient horizontal mounting of the larvae on a coverslip. A layer of 1% w/v agarose was laid on the coverslip, in which a 0.8 mm wide groove cast with a 3D-printed comb was created to hold the larvae during imaging, as described in ref. 30 Experiments were conducted with both objectives. High-resolution details of 3 dpf embryos were obtained with the 60× objective, whereas images of the full vascular system of a 5 dpf embryo were collected in a single sweep of the translation stage with the 20× objective. Images at 60× magnification were recorded at 20 Hz on the full 2048 by 1024 pixel active region, with a total data throughput of approximately 42×106  pixels/s, whereas 20× datasets were acquired at 50 Hz on a 2048 by 768 pixel active region, with a total data throughput of approximately 78×106  pixels/s. Three-dimensional datasets were captured by moving the translation stage, with horizontal spacing between frames of 0.4  μm for 60× images, and of 1  μm for 20× images.

Typical datasets are reported in Fig. 5. Images acquired with the 60× 1.1 NA objective showed fine details in the vascular structure, both in the relatively clear tail sections and in the more complex and optically dense vasculature of the eye. The 20× objective produced, as expected, lower brightness and lower resolution images, but its wide field of view enabled the acquisition of images of the entire 3.5 mm long larva in a single sweep of the actuator in approximately one minute. Image decorrelation analysis reported a lateral resolution of 3.3  μm for the 20× images, 0.9  μm for the 60X image in the tail, and 1.73  μm for the eye. These estimates are compatible with the resolution measured with microbeads, when considering the imperfect index match of the zebrafish embryos with water. Although the numerical aperture is lower, and therefore the quality of the images is not comparable, the field of view and pixel throughput of the microscope are comparable to state-of-the-art OPM with custom tertiary objectives.31

Fig. 5

Zebrafish larvae imaging. (a) and (b) Representative raw images from the detector during stack acquisition, with 60× and 20× objectives respectively. Scale bars are 50  μm in panel (a) and 100  μm in panel (b). (c) Typical two-dimensional lateral image after affine transform for a 60× image of the larva tail vasculature. Scale bar is 50  μm. (d) Typical two-dimensional lateral image after affine transformation for a 20× image of the larva vasculature. Scale bar is 50  μm. (e) and (f) Depth encoded projection of full dataset of the eye and tail vasculature respectively in 3 dpf larvae, acquired with the 60× objective at a camera frame rate of 20 Hz (total image time of 42 and 60 s, respectively), and scale bars are 100  μm. (g) Maximum intensity projection of an entire 5 dpf larva acquired with the 20× objective at a camera frame rate of 50 Hz (total image time of 64 s); scale bar is 200  μm.

JBO_29_3_036502_f005.png

4.

Discussion

This article introduces a novel approach to the visualization of tilted planes in OPM, employing an ETL and the detector rolling shutter to replace the reprojection setup of standard OPM. While the limited exposure time due to rolling shutter synchronization and the limited performance of the ETL prevent RF-OPM from reaching the volumetric frequencies of state of the art OPM implementations,32 the proposed setup presents several advantages compared to standard OPM, namely:

  • The system collects photons from the full aperture of the objective. While the gain in terms of effective signal is reduced by the need to have an exposure time shorter than the frame time to synchronize the rolling shutter with the ETL scan, this still enables imaging at higher resolution than OPM with cropped apertures.33 Similar performance can be achieved with discontinuities in the refractive index between the secondary and tertiary objectives of standard OPM, but the presented method is arguably simpler and cost-effective.

  • The proposed method allows great flexibility in the selection of the microscope objective. Unlike in standard OPM, low-aperture objectives can be used. Moreover, the objective can be rapidly switched during experiments, without the need to realign the optical system. The only limitation to the capability of operation with different objectives lies in the need to match the diameter of the back aperture of the objective with the diameter of the ETL. When the aperture diameter of the chosen objectives differs, as it did in the case of the ones employed here, the setup should be designed to either crop the pupil of the objectives with larger back apertures, or to underfill the ETL aperture when using objectives with smaller back apertures. Since, from Eq. (5), reducing the magnification of the objective pupil at the ETL plane results in smaller axial shifts for the same lens current input, the presented setup was designed to always use the full aperture of the ETL and maximize the axial field of view when using the 60× objective, cropping the aperture of the 20× objective as a tradeoff. Different experimental needs may require different tradeoffs. Ideally, alternative remote focusing approaches with faster actuators and longer ranges may allow the use of the full aperture of any objective needed.

  • The use of the same ETL for both excitation and fluorescence light ensures the light-sheet is always at its thinnest point in the region of the exposed rolling shutter. This greatly improves resolution at the axial edges of the field of view, and provides imaging of wider fields of view when compared to standard OPMs, or even traditional light-sheet microscopes with fixed cylindrical lenses. While this feature could be fully exploited in the presented data with the lower aperture objective, the narrow optimal axial range of ETL-based remote focusing with high aperture objectives does hinder this capability of the system, due to the appearance of non-negligible spherical aberration when focusing further from the objective’s focal plane. This position-dependent aberration could be reduced through the implementation of a more complex remote focusing method, or through the use of multi-conjugate aberration correction in the system.25

  • While the system reported in this article was realized to only operate with the shallowest possible light sheet angle achievable with a given objective, future implementations with a computer controlled tilt on mirror M2 from Fig. 2 would allow a vast range of adjustments, such as extension of the axial field of view, or multi-directional acquisition compensating for shadowing artifacts.31

Disclosures

The authors declare no conflicts of interest.

Code and Data Availability

All the raw data for the images presented in the manuscript, together with software for affine transformation are available as an open dataset on Zenodo [doi: https://doi.org/10.5281/ZENODO.10036881].

Acknowledgments

The research has received funding from LASERLAB-EUROPE (Grant No. 871124) (European Union’s Horizon 2020 research and innovation programme) and the European Union’s NextGenerationEU Programme with the I-PHOQS Infrastructure (Grant Nos. IR0000016, ID D2B8D520, and CUP B53C22001750006) “Integrated infrastructure initiative in Photonic and Quantum Sciences.” The PhD student Alessia Brix was supported by the PhD program in Experimental Medicine of the University of Milan, Italy.

References

1. 

J. M. Girkin and M. T. Carvalho, “The light-sheet microscopy revolution,” J. Opt., 20 (5), 053002 https://doi.org/10.1088/2040-8986/aab58a (2018). Google Scholar

2. 

A. Candeo et al., “Light sheet fluorescence microscopy quantifies calcium oscillations in root hairs of arabidopsis thaliana,” Plant Cell Physiol., 58 (7), 1161 –1172 https://doi.org/10.1093/pcp/pcx045 PCPHA5 0032-0781 (2017). Google Scholar

3. 

A. Candeo et al., “Virtual unfolding of light sheet fluorescence microscopy dataset for quantitative analysis of the mouse intestine,” J. Biomed. Opt., 21 (5), 056001 https://doi.org/10.1117/1.JBO.21.5.056001 JBOPFO 1083-3668 (2016). Google Scholar

4. 

A. K. Glaser et al., “A hybrid open-top light-sheet microscope for versatile multi-scale imaging of cleared tissues,” Nat. Methods, 19 (5), 613 –619 https://doi.org/10.1038/s41592-022-01468-5 1548-7091 (2022). Google Scholar

5. 

C. Dunsby, “Optically sectioned imaging by oblique plane microscopy,” Opt. Express, 16 (25), 20306 –20316 https://doi.org/10.1364/OE.16.020306 OPEXFF 1094-4087 (2008). Google Scholar

6. 

J. Kim, “Recent advances in oblique plane microscopy,” Nanophotonics, 12 (13), 2317 –2334 https://doi.org/10.1515/nanoph-2023-0002 (2023). Google Scholar

7. 

B. Yang et al., “Epi-illumination spim for volumetric imaging with high spatial-temporal resolution,” Nat. Methods, 16 (6), 501 –504 https://doi.org/10.1038/s41592-019-0401-3 1548-7091 (2019). Google Scholar

8. 

E. Sapoznik et al., “A versatile oblique plane microscope for large-scale and high-resolution imaging of subcellular dynamics,” Elife, 9 e57681 https://doi.org/10.7554/eLife.57681 (2020). Google Scholar

9. 

M. Hoffmann and B. Judkewitz, “Diffractive oblique plane microscopy,” Optica, 6 (9), 1166 –1170 https://doi.org/10.1364/OPTICA.6.001166 (2019). Google Scholar

10. 

F. O. Fahrbach et al., “Rapid 3d light-sheet microscopy with a tunable lens,” Opt. Express, 21 (18), 21010 –21026 https://doi.org/10.1364/OE.21.021010 OPEXFF 1094-4087 (2013). Google Scholar

11. 

E. Baumgart and U. Kubitscheck, “Scanned light sheet microscopy with confocal slit detection,” Opt. Express, 20 (19), 21805 –21814 https://doi.org/10.1364/OE.20.021805 OPEXFF 1094-4087 (2012). Google Scholar

12. 

K. M. Dean et al., “Isotropic imaging across spatial scales with axially swept light-sheet microscopy,” Nat. Protoc., 17 (9), 2025 –2053 https://doi.org/10.1038/s41596-022-00706-6 1754-2189 (2022). Google Scholar

13. 

H. R. Ueda et al., “Whole-brain profiling of cells and circuits in mammals by tissue clearing and light-sheet microscopy,” Neuron, 106 (3), 369 –387 https://doi.org/10.1016/j.neuron.2020.03.004 NERNET 0896-6273 (2020). Google Scholar

14. 

F. Ströhl et al., “Multifocus microscopy with optical sectioning and high axial resolution,” Optica, 9 (11), 1210 –1218 https://doi.org/10.1364/OPTICA.468583 (2022). Google Scholar

15. 

S.-T. Hung et al., “Soleil: single-objective lens inclined light sheet localization microscopy,” Biomed. Opt. Express, 13 (6), 3275 –3294 https://doi.org/10.1364/BOE.451634 BOEICL 2156-7085 (2022). Google Scholar

16. 

J. Jiang et al., “Fast 3-D temporal focusing microscopy using an electrically tunable lens,” Opt. Express, 23 (19), 24362 –24368 https://doi.org/10.1364/OE.23.024362 OPEXFF 1094-4087 (2015). Google Scholar

17. 

T. Wright et al., “Video-rate remote refocusing through continuous oscillation of a membrane deformable mirror,” J. Phys.: Photonics, 3 (4), 045004 https://doi.org/10.1088/2515-7647/ac29a2 (2021). Google Scholar

18. 

M. Bawart et al., “Remote focusing in confocal microscopy by means of a modified alvarez lens,” J. Microsc., 271 (3), 337 –344 https://doi.org/10.1111/jmi.12724 JMICAR 0022-2720 (2018). Google Scholar

19. 

E. J. Botcherby et al., “An optical technique for remote focusing in microscopy,” Opt. Commun., 281 (4), 880 –887 https://doi.org/10.1016/j.optcom.2007.10.007 OPCOB8 0030-4018 (2008). Google Scholar

20. 

E. J. Botcherby et al., “Aberration-free optical refocusing in high numerical aperture microscopy,” Opt. Lett., 32 (14), 2007 –2009 https://doi.org/10.1364/OL.32.002007 OPLEDP 0146-9592 (2007). Google Scholar

21. 

A. K. Chmielewski et al., “Fast imaging of live organisms with sculpted light sheets,” Sci. Rep., 5 (1), 9385 https://doi.org/10.1038/srep09385 SRCEC3 2045-2322 (2015). Google Scholar

22. 

M. Mickoleit et al., “High-resolution reconstruction of the beating zebrafish heart,” Nat. Methods, 11 (9), 919 –922 https://doi.org/10.1038/nmeth.3037 1548-7091 (2014). Google Scholar

23. 

D. P. Ryan et al., “Automatic and adaptive heterogeneous refractive index compensation for light-sheet microscopy,” Nat. Commun., 8 (1), 612 https://doi.org/10.1038/s41467-017-00514-7 NCAOBW 2041-1723 (2017). Google Scholar

24. 

J. Kim et al., “Vectorial point spread function and optical transfer function in oblique plane imaging,” Opt. Express, 22 (9), 11140 –11151 https://doi.org/10.1364/OE.22.011140 OPEXFF 1094-4087 (2014). Google Scholar

25. 

T. Furieri, A. Bassi and S. Bonora, “Large field of view aberrations correction with deformable lenses and multi conjugate adaptive optics,” J. Biophotonics, 16 (12), e202300104 https://doi.org/10.1002/jbio.202300104 (2023). Google Scholar

26. 

D. Wilding et al., “Adaptive illumination based on direct wavefront sensing in a light-sheet fluorescence microscope,” Opt. Express, 24 (22), 24896 –24906 https://doi.org/10.1364/OE.24.024896 OPEXFF 1094-4087 (2016). Google Scholar

27. 

A. Descloux, K. S. Grußmayer and A. Radenovic, “Parameter-free image resolution estimation based on decorrelation analysis,” Nat. Methods, 16 (9), 918 –924 https://doi.org/10.1038/s41592-019-0515-7 1548-7091 (2019). Google Scholar

28. 

S.-W. Jin et al., “Cellular and molecular analyses of vascular tube and lumen formation in zebrafish,” Development, 132 (23), 5199 –5209 https://doi.org/10.1242/dev.02087 (2005). Google Scholar

29. 

J. Karlsson, J. Von Hofsten and P.-E. Olsson, “Generating transparent zebrafish: a refined method to improve detection of gene expression during embryonic development,” Mar. Biotechnol., 3 522 –527 https://doi.org/10.1007/s1012601-0053-4 (2001). Google Scholar

30. 

A. R. Ahmed et al., “Directly imaging the localisation and photosensitization properties of the pan-mtor inhibitor, azd2014, in living cancer cells,” J. Photochem. Photobiol. B: Biol., 213 112055 https://doi.org/10.1016/j.jphotobiol.2020.112055 JPPBEG 1011-1344 (2020). Google Scholar

31. 

B. Yang et al., “Daxi—high-resolution, large imaging volume and multi-view single-objective light-sheet microscopy,” Nat. Methods, 19 (4), 461 –469 https://doi.org/10.1038/s41592-022-01417-2 1548-7091 (2022). Google Scholar

32. 

S. Daetwyler et al., “Mesoscopic oblique plane microscopy via light-sheet mirroring,” Optica, 10 (11), 1571 –1581 https://doi.org/10.1364/OPTICA.502243 (2023). Google Scholar

33. 

V. Voleti et al., “Real-time volumetric microscopy of in vivo dynamics and large-scale samples with scape 2.0,” Nat. Methods, 16 (10), 1054 –1062 https://doi.org/10.1038/s41592-019-0579-4 1548-7091 (2019). Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Paolo Pozzi, Vipin Balan, Alessia Candeo, Alessia Brix, Anna Silvia Pistocchi, Cosimo D’Andrea, Gianluca Valentini, and Andrea Bassi "Full-aperture extended-depth oblique plane microscopy through dynamic remote focusing," Journal of Biomedical Optics 29(3), 036502 (21 March 2024). https://doi.org/10.1117/1.JBO.29.3.036502