Paper
21 August 2013 Accurate focusing of knife-edge imaging optical system based on the area under MTF curve
Bin Wu, Li-gong Li, Shang-qian Liu, Cheng-ping Ying, Kun-feng Chen
Author Affiliations +
Proceedings Volume 8908, International Symposium on Photoelectronic Detection and Imaging 2013: Imaging Sensors and Applications; 89081U (2013) https://doi.org/10.1117/12.2034545
Event: ISPDI 2013 - Fifth International Symposium on Photoelectronic Detection and Imaging, 2013, Beijing, China
Abstract
The method of knife-edge scanning is always adopted in testing the MTF of infrared focal plane array (IRFPA), and accurate focusing is one of the most important preconditions in the measurement procedure, because the measurement accuracy of MTF is ensured only when IRFPA is placed in the focal plane of the knife-edge imaging optical system. In this paper, a focusing method based on the area value under normalized MTF curve is proposed. Firstly, we analyzed MTF calculation model, which contained derivation of edge spread function (ESF) and then Fourier transformation. In this way, the issue of larger area under normalized MTF curve meant better focusing degree of the measurement system was proved. Next, the quasi-focal plane position of knife-edge optical system was determined according to the output voltage values of pixels of IRFPA. Then a series of MTF curves were measured when IRFPA was placed at different positions located in the optical axis beside the determined quasi-focal plane at an increment of 60 μm. Subsequently, the area under each normalized MTF curve was calculated, and the result showed the area values presented a perfect Lorentzian distribution against the positions of IRFPA. It was concluded the peak position of the fitted Lorentzian curve corresponded to the position of focal plane of knife-edge optical system. With this method, the focal plane position of knife-edge optical system was determined. The method presented here can be used to focus the MTF testing system with a high accuracy, which is good for the following MTF measurement.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bin Wu, Li-gong Li, Shang-qian Liu, Cheng-ping Ying, and Kun-feng Chen "Accurate focusing of knife-edge imaging optical system based on the area under MTF curve", Proc. SPIE 8908, International Symposium on Photoelectronic Detection and Imaging 2013: Imaging Sensors and Applications, 89081U (21 August 2013); https://doi.org/10.1117/12.2034545
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation transfer functions

Imaging systems

Optical imaging

Infrared radiation

Infrared imaging

Reflectors

Spatial frequencies

RELATED CONTENT

Method of object MRTD-testing for thermal infrared imager
Proceedings of SPIE (November 05 2018)
Assessing The Performance Of Thermal Imagers
Proceedings of SPIE (October 11 1988)
MTF measurement of infrared optical systems
Proceedings of SPIE (September 23 2009)
Measurement of the MTF of IR staring-array imaging systems
Proceedings of SPIE (September 16 1992)
Blackbody source technology trends
Proceedings of SPIE (May 09 2011)

Back to Top