Paper
11 October 2004 Characterizing nonlinearities in the Chandra LETG+HRC-S dispersion relation
Author Affiliations +
Abstract
The Chandra Low Energy Transmission Grating Spectrometer (LETGS) is comprised of 3 micro-channel plate (MCP) segments and is primarily used with the High Resolution Camera spectroscopic array (HRC-S). In-flight calibration data observed with the LETG+HRC-S show that there are non-linear deviations in the positions of some lines by as much as 0.1 Å. These deviations are thought to be caused by spatial non-linearities in the imaging characteristics of the HRC-S detector. Here, we present the methods we used to characterize the non-linearities of the dispersion relation across the central plate of the HRC-S, and empirical corrections which greatly reduce the observed non-linearities by a factor of 2 or more on the central MCP.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sun Mi Chung, Jeremy J. Drake, Vinay L. Kashyap, Peter W. Ratzlaff, and Bradford J. Wargelin "Characterizing nonlinearities in the Chandra LETG+HRC-S dispersion relation", Proc. SPIE 5488, UV and Gamma-Ray Space Telescope Systems, (11 October 2004); https://doi.org/10.1117/12.552194