Paper
19 November 1998 Optimizing the ACIS effective area and energy resolution
George Chartas, Gordon P. Garmire, John A. Nousek, Scott Koch, Steven E. Kissel, Gregory Y. Prigozhin, Mark W. Bautz
Author Affiliations +
Abstract
The ACIS flight instrument was recently extensively calibrated at the X-ray Calibration Facility at MSFC. For the analysis of a subset of the ACIS calibration data we have employed an automated even filtering software package developed at PSU. We present result describing the dependence of the Effective Area and Energy Response of ACIS/HRMA as a function of grade selection, split event threshold, CCD and CCD amplifier, and off-axis angle. The main goal of this study is to facilitate the selection of the appropriate ACIS parameters that optimize a desired feature in an observed ACIS spectrum and may also guide observers in selecting the appropriate CCD for the observation. Optimizing a particular feature in an ACIS observation in general may require making a trade-off between effect area and energy response of AXAF/ACIS. The present analysis will facilitate the selection of the appropriate grades, CCD, CCD amplifier and spilt even thresholds needed to attain the optimal point.As an illustration of the effectiveness of this approach we present several case studies of typical astrophysical source spectra in which we enhance a particular scientific feature in the observed ACIS spectrum by the appropriate selection of ACIS parameters.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
George Chartas, Gordon P. Garmire, John A. Nousek, Scott Koch, Steven E. Kissel, Gregory Y. Prigozhin, and Mark W. Bautz "Optimizing the ACIS effective area and energy resolution", Proc. SPIE 3444, X-Ray Optics, Instruments, and Missions, (19 November 1998); https://doi.org/10.1117/12.331241
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Charge-coupled devices

X-rays

Calibration

Amplifiers

Spectral resolution

X-ray detectors

Fourier transforms

RELATED CONTENT

X-ray detectors and calibration system for the MAXI mission
Proceedings of SPIE (October 22 1999)
PN CCD camera for XMM performance of full frame...
Proceedings of SPIE (October 15 1997)
Using tritium and x ray tubes as x ray calibration...
Proceedings of SPIE (October 31 1996)
ASTRO-E/XRS calibration program and results
Proceedings of SPIE (October 22 1999)
Flight CCD selection for the x ray CCD camera of...
Proceedings of SPIE (February 03 2004)

Back to Top