28 March 2024 The energy response of extended CT ionization chamber was studied by MCNP program and verified by experiment
Author Affiliations +
Proceedings Volume 13091, Fifteenth International Conference on Signal Processing Systems (ICSPS 2023); 130912K (2024) https://doi.org/10.1117/12.3024960
Event: Fifteenth International Conference on Signal Processing Systems (ICSPS 2023), 2023, Xi’an, China
Abstract
At present, the effective length of CT ionization chamber commonly used in China is 100mm. When the nominal line is wider than 60mm, the measurement accuracy decreases, and multiple step measurements are required, which is troublesome to operate. In this paper, the energy response of the extended CT ionization chamber with different materials is studied. The MCNP5 program is simulated and verified by the standard radiation field energy response experiment. The experimental results show that the energy response of the collector electrode with 1mm Al core coated with 0.1mm graphite is ±2% in the energy range of CT (100-150) kV, which has good application value.
(2024) Published by SPIE. Downloading of the abstract is permitted for personal use only.
Lan Gong, HongYou Yin, Wei Liu, WenLong Dong, ZhiXin Zheng, and Chuan Wei "The energy response of extended CT ionization chamber was studied by MCNP program and verified by experiment", Proc. SPIE 13091, Fifteenth International Conference on Signal Processing Systems (ICSPS 2023), 130912K (28 March 2024); https://doi.org/10.1117/12.3024960
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Vacuum chambers

Ionization

X-ray computed tomography

Aluminum

Electrodes

X-rays

Electrons

Back to Top