Paper
13 July 2022 Evaluation of 3D printed contrast detail phantoms for mammography quality assurance
Author Affiliations +
Proceedings Volume 12286, 16th International Workshop on Breast Imaging (IWBI2022); 122860J (2022) https://doi.org/10.1117/12.2625732
Event: Sixteenth International Workshop on Breast Imaging, 2022, Leuven, Belgium
Abstract
Objects created by 3D printers are increasingly used in various medical applications. Today, affordable 3D printers, using Fused Deposition Modeling are widely available. In this project, a commercially available 3D printer was used to replicate a conventional radiographic contrast detail phantom. Printing materials were selected by comparing their x-ray attenuation properties. Two replicas were printed using polylactic acid, with different filling patterns. The printed phantoms were imaged by a clinical mammography system, using automatic exposure control. Phantom images were visually and quantitively compared to images of the corresponding conventional contrast detail phantom. Visual scoring of the contrast detail elements was performed by a medical physics student. Contrast-to-noise ratio (CNR) was calculated for each phantom element. The diameter and thickness of the smallest visible phantom object were 0.44 mm and 0.09 mm, respectively, for both filling patterns. For the conventional phantom, the diameter and thickness of the smallest visible object were 0.31 mm and 0.09 mm. Visual inspection of printed phantoms revealed some linear artefacts. These artefacts were however not visible on mammographic projections. Quantitively, average CNR of printed phantom objects followed the same trend with an increase of average CNR with increasing disk height. However, there is a limitation of detail objects with disk diameters below 1.25 mm, caused by the available nozzle size. Based upon the encouraging results, future work will explore the use of different materials and smaller nozzle diameters.
© (2022) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Måns Boll, Trevor Vent, Hanna Tomic, Christian Bernhardsson, Magnus Dustler, Anders Tingberg, and Predrag R. Bakic "Evaluation of 3D printed contrast detail phantoms for mammography quality assurance", Proc. SPIE 12286, 16th International Workshop on Breast Imaging (IWBI2022), 122860J (13 July 2022); https://doi.org/10.1117/12.2625732
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
3D printing

Fused deposition modeling

Polymethylmethacrylate

Mammography

Back to Top