We proposed a direct-indirect dual layer detector combination for CEDM and CEDBT to eliminate patient motion artifacts in dual energy images. Both physical experiments and Monte Carlo simulations were conducted to compare dual shot technique with the dual layer technique. The proposed direct-indirect dual layer detector combination uses a 200 µm direct a-Se as the front detector for LE images and a 400~1000 µm indirect CsI as the back detector for HE images. Results showed similar breast tissue cancellation between dual layer and dual shot techniques. Dual layer technique with direct-indirect detector has higher SDNR/MGD compared with dual shot technique.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The alert did not successfully save. Please try again later.
Xiaoyu Duan, Adrian Howansky, Hailiang Huang, Wei Zhao, "Evaluation of a direct-indirect dual layer detector for contrast enhanced breast imaging: Monte Carlo simulations and physical experiments," Proc. SPIE 12031, Medical Imaging 2022: Physics of Medical Imaging, 120310A (4 April 2022); https://doi.org/10.1117/12.2613056