Paper
28 December 1989 Results Of Aynamic Range Measurements On Muitianoae Microchannel Array Detectors
Vic S. Argabright, Richard L. Bybee
Author Affiliations +
Abstract
We will present results showing the single pixel counting linearity of the MAMA (MultiAnode Microchannel Array) detector system for three different tubes. The relationship of MCP (microchannel plate) gain characteristics, amplifier discrimination level and ion feedback to single pixel counting linearity will be discussed. Comparisons of single pixel linearity between tubes using curved MCF's and a chevron pair will also be discussed.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Vic S. Argabright and Richard L. Bybee "Results Of Aynamic Range Measurements On Muitianoae Microchannel Array Detectors", Proc. SPIE 1158, Ultraviolet Technology III, (28 December 1989); https://doi.org/10.1117/12.962537