The High Resolution Energetic X-ray Imager (HREXI ) is a coded-aperture imaging telescope that utilizes a large closely-tiled array of CdZnTe (CZT) detectors, each 19.9 x 19.9 x 3mm with a 32 x 32 pixel (604μm) for coded aperture X-ray imaging (3 - 200 keV) of cosmic X-ray sources and transients. Each CZT crystal is read out by an ASIC incorporating, for the first time, Through Silicon Vias (TSVs). These TSVs replace the wire bonds for this ASIC, originally designed for the Nuclear Spectroscopic Telescope Array (NuSTAR) focusing hard X-ray telescope. The TSVs allow flip-chip bonding of the ASIC to the PCB board electronics for processing of the data. The new TSV-ASICs will enable closer tiling and larger imaging arrays which require faster, more efficient ASIC testing and calibration at the die level. We have designed and developed an ASIC Test Stand (ATS) for rapid ASIC testing prior to bonding to CZT. We demonstrate how ASIC die-level testing with the ATS can be performed rapidly with rigidly spaced micro-pogo pins supported by an FPGA readout.
|