18 April 2023 Prediction of motion artifacts caused by translation in handheld laser speckle contrast imaging
Author Affiliations +
Abstract

Significance

In handheld laser speckle contrast imaging (LSCI), motion artifacts (MA) are inevitable. Suppression of MA leads to a valid and objective assessment of tissue perfusion in a wide range of medical applications including dermatology and burns. Our study shines light on the sources of these artifacts, which have not yet been explored. We propose a model based on optical Doppler effect to predict speckle contrast drop as an indication of MA.

Aim

We aim to theoretically model MA when an LSCI system measuring on static scattering media is subject to translational displacements. We validate the model using both simulation and experiments. This is the crucial first step toward creating robustness against MA.

Approach

Our model calculates optical Doppler shifts in order to predict intensity correlation function and contrast of the time-integrated intensity as functions of applied speed based on illumination and detection wavevectors. To validate the theoretical predictions, computer simulation of the dynamic speckles has been carried out. Then experiments are performed by both high-speed and low-framerate imaging. The employed samples for the experiments are a highly scattering matte surface and a Delrin plate of finite scattering level in which volume scattering occurs.

Results

An agreement has been found between theoretical prediction, simulation, and experimental results of both intensity correlation functions and speckle contrast. Coefficients in the proposed model have been linked to the physical parameters according to the experimental setups.

Conclusions

The proposed model provides a quantitative description of the influence of the types of illumination and media in the creation of MA. The accurate prediction of MA caused by translation based on Doppler shifts makes our model suitable to study the influence of rotation. Also the model can be extended for the case of dynamic media, such as live tissue.

1.

Introduction

Four decades ago, a technique was developed for the measurement of spatiotemporal blood flow variations in the microcirculation by the analysis of laser speckle.1 Since then, this technique, known as laser speckle contrast imaging (LSCI),25 has advanced in different aspects, such as enhancement of spatial resolution,6 processing time,7 and quantitative prediction of blood flow.811 When coherent light encounters an optically rough surface or propagates through a scattering medium such as tissue, an interference pattern called speckle is observed on an image screen. Movements of the medium or scattering particles inside it cause formation of dynamic speckles. Imaging of dynamic speckles with a limited exposure time results in speckle pattern images of reduced contrast. The speckle contrast is a measure of motion on or inside the illuminated target: lower speckle contrast corresponds to higher movement.3

LSCI covers a wide range of clinical applications,12 including psoriasis,13,14 diabetic foot,15 and blood flow monitoring during surgery (intraoperative).16,17 Operated in handheld mode, LSCI brings convenience for both patients and clinical staff.18 However, involuntarily movements of the subject or the operator result in movement artifact, that is spatial shifting of the consecutive images (off-axis imaging) as well as extra blurring of individual speckle patterns. The former can be overcome by movement detection and image registration.19 The latter is more complicated and has been addressed so far in different ways: (i) motion stabilization using a gimbal mount as a prevention of motion artifacts (MA) due to operator motion20 and (ii) capturing speckle patterns from an opaque surface mounted in the field of view during the measurements. The perfusion measured on the opaque surface was used either to choose a number of movement-artifact-free frames21 or to be removed from the perfusion values measured on tissue as a compensation of MA.2225 Altogether, a fundamental model to explain the underlying reason but also to predict MA is missing.

In a previous work, we studied MA, defined by the drop in speckle contrast due to the relative movements of the device and the object of measurement, due to the translation, rotation, and tilt of wavefronts. We found that the lower the scattering level of the medium is, the higher the drop of the speckle contrast is.26 In another study, we showed that use of planar or spherical waves for illumination instead of conventional scrambled waves made by glass (engineered) diffusers is a hardware-based and straightforward approach to decrease the MA.27

Then we carried out a comparison between handheld and mounted measurements on psoriasis patients. Results suggested that although the pair of mounted and handheld measurements were in agreement on a visual basis, the difference in lesion perfusion between mounted and handheld measurements was on-average statistically significant.28 This statistically significant difference caused by the movement-induced perfusion in handheld measurements may limit certain applications of handheld LSCI.

In this work, we use the concept of optical Doppler effect29 to predict speckle contrast drop due to translation and for solid objects. This is a first step in modeling MA due to linear motion and rotation when measuring on objects with internal motion, such as perfused tissue. This paper presents a theoretical model that describes part of the MA, namely translation on a highly scattering surface (matte paint) and a tissue-level scattering object (Delrin) validated by simulation of dynamic speckles and experiments.

2.

Model of the Optical System and Speckle Contrast Generation

2.1.

Optical Doppler Shifts Induced by Translation of a Solid Object

Consider a solid object of finite scattering level moving with a constant velocity v along the xy plane in the x direction [see Fig. 1(a)]. The object is illuminated over a large area by a coherent light source. For a given point A on the object’s surface, a cone of base radius ri is formed that is defined by the range of incoming wavevectors ki. Assuming a lens collecting scattered light at point B located at the object’s surface, a cone of base radius rs defined by the range of wavevectors ks will be formed. The collecting lens will conjugate the point B to a point on the imaging screen of a detector causing time varying intensity I(t) containing a range of frequencies. The core of our model is that these frequencies are assumed to be dictated by the optical Doppler effect as29

Eq. (1)

ωD=v.(kski).

Fig. 1

Schematic diagram of the system model reprinted with permission from Ref. 30. Linear translation of a solid object along the xy plane in the x direction. Illumination and detection on a solid object of (a) finite and (b) very high scattering levels. (c) Mapping of the bases of the illumination and detection cones on the object’s surface (xy plane).

JBO_28_4_046005_f001.png

The type of illumination determines the spread of ki and, therefore, the size of ri. For scrambled waves made by a diffuser, the ri takes a nonzero value. For spherical waves or plane waves, there is only a single ki on A, which leads ri to be zero. Notice that for spherical waves, the single ki is position dependent and for scrambled waves, the orientation of the cone of ki is position dependent. The openness of a diaphragm located at the base of a detection cone controls the spread of ks and consequently the size of rs. If the scattering medium is of finite scattering level, then for each detection point B on the object’s surface, the scattered light originally entering the medium at a range of positions around B contributes to the observed Doppler shift introduced in Eq. (1). This range depends on the type of illumination and the optical properties of the medium. Note that the relative speed between the pair of illumination-detection and scattering object at each moment forms v in Eq. (1); thus either of the two can be subjected to translational movements.

2.1.1.

Highly scattering medium

Consider the scattering medium shown in Fig. 1(a) to be of high scattering level such that no light penetrates into the medium (e.g., a white matte surface). Then A and B will coincide, and for each observation point B, only the incoming wavevectors to B and the outgoing wavevectors from B will contribute to creation of Doppler shifts on the conjugated point on the detector. Now the problem narrows down to a simplified geometry as shown in Fig. 1(b). If the bases of illumination and detection cones are mapped on the xy plane and assuming that the angles between the normal to the surface and both ki and ks are relatively small, then the begin points of ki(x,y) and the end points of all ks(x,y) occupy a circular region with radius ri   and rs, respectively, as shown in Fig. 1(c). Since the object moves in the x direction, the consequence of Eq. (1) is that all the wavevectors differences indicated by Δx=x2x1 [see Fig. 1(c)] will create the same Doppler shift. The intensity fluctuations are caused by mixing light at various Doppler shifts, whereas light with a single Doppler shift does not yield an intensity fluctuation by itself. The frequencies of the fluctuations are caused by the variation of the Doppler shifts but not the absolute Doppler shifts. Consequently, the absolute tilt angles of the wavevectors are not of importance but the wavevector spreads that are modeled by ri and rs.

From Eq. (1), a single Doppler shift as a function of the wavevector difference can be written as

Eq. (2)

ωD(Δx)=2πVλ[ls(x2),li(x1)],
where V is absolute speed, λ is the wavelength of the light source, and ls(.) and li(.)=li(.) are the unit vector elements that depend on the spreads of ks(x,y) and ki(x,y), respectively. These vector elements allow us to model the dependency of each single Doppler shift in Eq. (2) to the x-related distance between any two points in the circles shown in Fig. 1(c). The higher the spreads of k vectors, the larger the radii of these circles. A Doppler distribution is made by adding all the single Doppler shifts introduced in Eq. (2) as

Eq. (3)

A(ωDλ2πV)=pRi(ωDλ2πV)*pRs(ωDλ2πV),
where * denotes the convolution operator, and pRi(.) and pRs(.) are the probability density functions of the spreads of incoming and outgoing wavevectors, respectively. Thus the right-hand side of Eq. (3) represents an exact mathematical expression of individual wavevector differences lsi(.) introduced in Eq. (2). The power spectral density can be calculated as31

Eq. (4)

P(ωDλ2πV)=A(ωDλ2πV)*A(ωDλ2πV).

Then the normalized intensity autocorrelation is calculated by

Eq. (5)

g(2)(τ)=F1{P(ωDλ2πV)}  ,
where F1{.} is the inverse Fourier transformation operator. Equation (5) is a representation of the Wiener–Khinchin theorem.32 Now, based on the properties of the Fourier transform,33 g(2)(τ) can be calculated from pRi and pRs as

Eq. (6)

g(2)(τ)=[F1{pRi}]2[F1{pRs}]2.

The temporal contrast of time-integrated speckle intensity is

Eq. (7)

C2=2T0T(1τT)|g(1)(τ)|2dτ,
where g(1)(τ)=g(2)(τ)1 is the normalized complex speckle field correlation34 (Siegert relation). See Sec. 1 in the Supplementary Material for the proof of Eq. (7) that also holds for spatial contrast of time-integrated intensity.35 Therefore, on a highly scattering object (i.e., infinite scattering coefficient), it is possible to predict the speckle contrast drop as a result of applied v when there is knowledge about pRi and pRs. The probability density functions pRi and pRs can take several possible forms of which we consider two, namely, semicircular and Gaussian.

Semicircular density function of the wavevectors

We define the following form for a semicircular form of the density functions:

Eq. (8)

pi,s(ωDλ2πV)=λπ2Vχi,s1(ωDλ2πVχi,s)2,
where χi and χs are the shape parameters that control the spread of pRi and pRs, respectively. In Fig. 2(a), a pair of pRi and pRs with semicircular shapes is represented. The corresponding optical Doppler histogram is shown in Fig. 2(b). Substituting Eq. (8) into Eq. (6) and performing the calculation results in

Eq. (9)

g(2)(τ)=16bi2bs2(J1(τbi)J1(τbs)τ2)2+1,
where J1(.) is the Bessel function of the first kind36 and bi,s=λ/(2πVχi,s). By substituting the calculated g(1)(τ) from Eq. (9) into Eq. (7), the contrast of the time-integrated intensity takes the following form:

Eq. (10)

C2=32bibsT0T1τ4J1(τbi)2J1(τbs)2[1τT]dτ.

Fig. 2

Density functions for incoming and outgoing wavevectors. λ=671  nm; V=5  mm/s. (a) Case of a semicircular shape and (b) the associated optical Doppler distribution as a result of convolution of pRi and pRs. (c) Case of a Gaussian shape and (d) the associated optical Doppler distribution.

JBO_28_4_046005_f002.png

Equation (10) cannot be solved analytically; however, it can be calculated numerically.

Gaussian density function of the wavevectors

The Gaussian form of pRi and pRs can be written as follows:

Eq. (11)

pi,s(ωDλ2πV)=exp((ωDλ2πVσi,s)2),
where σi and σs are the shape parameters that control the spread of pRi and pRs, respectively. An example of a pair of pRi and pRs with Gaussian shapes and the corresponding optical Doppler histogram are shown in Figs. 2(c)2(d), respectively. Similar to the semicircular case, substitution of Eq. (11) into Eq. (6) results in

Eq. (12)

g(2)(τ)=exp(12aτ2)+1,
where a=(2πV/λ)2[σi2+σs2]. Finally, the substitution of the calculated g(1)(τ) from Eq. (12) into Eq. (7) results the following contrast of the time-integrated intensity:

Eq. (13)

C2=2T[π2aerf(Ta2)+exp(12aT2)1at],
where erf(z)=0zexp(t2)dt is the error function.

2.1.2.

Finite scattering medium

For a medium with finite scattering level, Eqs. (10) and (13) only hold for plane wave illumination. This is because for a finite scattering level, a single detection point B [Fig. 1(a)] is associated with a range of light injection positions A. Only for plane wave illumination, the illumination wavevector will be the same for all positions of injection of the light. Basically, with plane wave illumination, the observed Doppler shift will be independent of the light path ways from an arbitrary point A to the point B.

For the case of spherical waves, there is a single ki per arbitrary point A, but the direction of each ki depends on the location of A. Therefore, when illuminating with spherical or scrambled waves, information about the range of light injection positions that contribute to the detection in the observation point B is required. This information can be obtained by running a Monte-Carlo simulation of light propagation. This is beyond the scope of this paper.

2.2.

Simulation of Time-Varying Speckle Patterns of Size 1 Pixel

For validation of the proposed theory, a time-dependent speckle complex field is generated for each pixel at location (x,y) as

Eq. (14)

E(x,y,t)=1Nn=1Nexp(i(ωn(x,y)t+φn(x,y))),
where N is the number of phasors set as 50 in the simulation, j=1, ϕn(x,y) is a uniformly distributed random number on the interval [π,π] at each position (x,y), and ωn(x,y) is an angular frequency picked from the optical Doppler density function A(ωDλ/(2πV)) introduced in Eq. (3) in a weighted manner: the higher the amplitude of frequency in A(ωDλ/(2πV)), the higher the chance of selecting that frequency as ωn(x,y). In the dynamic speckle pattern obtained in this manner, each speckle occupies 1 pixel. Figure S1(a) in the Supplementary Material shows a simulated frame of the absolute field of size 100×100  px. The corresponding speckle intensity is calculated as [see Sec. 3 and Fig. S1(b) in the Supplementary Material]

Eq. (15)

I(x,y,t)=|E(x,y,t)|2.

The time-integrated speckle intensity is

Eq. (16)

IT(x,y)=0TI(x,y,t)dt,
where T denotes the exposure time. Finally, the spatial contrast of the time-integrated speckle intensity is calculated as

Eq. (17)

C=σIT(x,y)IT¯(x,y),
where σIT(x,y) and IT¯(x,y) represent the spatial standard deviation and the mean of IT(x,y), respectively.

2.3.

Imaging Geometry

We consider the diffraction of a single-lens imaging system in order to generate speckle patterns that simulate those obtained during experiments. To be specific, a value for the shape parameters χ and σ [introduced in Eqs. (8) and (11), respectively] based on the experimental setup is determined in order to simulate dynamic speckle patterns. This way, the proposed theory is evaluated using both simulation and experiments. In this study, the scattered light is collected by a single-lens imaging system consisting of an aperture, a spherical lens, and an image screen (see Fig. 3). The surface of the scattering medium is located at the object plane. The aperture and the spherical lens are located at the aperture plane. The image screen is located at the image plane. See Sec. 2 in the Supplementary Material for description of the transfer functions.

Fig. 3

Schematic drawing of the single-lens imaging system. S, scattering medium; A, aperture; L, spherical lens; I, image screen; z1, distance from the scattering medium to the lens; z2, distance from the lens to the image screen; and as, radius of the aperture opening.

JBO_28_4_046005_f003.png

A representation of simulated speckles in an imaging system with different aperture sizes for a time interval of 10 ms is provided in Figs. 4(a)4(d) in which the wavevector density functions are circular (see Sec. 3.2 for the distances and Sec. 4 for the values of χi and χs) and the applied translational speed is V=10  mm/s. These speckles are first generated pixel-by-pixel based on Eq. (14). Then the transfer functions described in Sec. 2 in the Supplementary Material are applied to them. Note that (S18) and the first term of (S19) in the Supplementary Material only apply phase shift to the speckle fields and the second term of (S19) in the Supplementary Material controls the speckle size. Therefore, it is also possible to directly generate them on the image screen and apply the spatial Fourier filter to account for the pupil function. This is equivalent to a single-lens imaging system that is considered for generation of speckles in the simulation part of this study. As a comparison with experimental speckle intensities, see Figs. 4(e) and 4(f), in which the speckle patterns obtained on a matte surface and a Delrin plate is shown, respectively. The experimental parameters for these two patterns are provided in Sec. 3.2. The speckle diameter obtained by spatial autocorrelation of intensities for Figs. 4(e) and 4(f) is about 3 pixels (each image pixel represents 63  μm), which is closest to the diameter of simulated speckles shown in Fig. 4(b), namely around 2 pixels. Note that the goal of speckle simulations is to predict temporal fluctuations of the speckles obtained via experiments rather than the speckle size. This is because it is the temporal fluctuations of the speckles that eventually govern the contrast of time-integrated intensity but not the speckle size. And since the opening of the pupil and the distance from the surface to the imaging system have been considered in the wavevector density function of the incoming waves, speckles of size 1 pixel contain all the optical Doppler shifts introduced in Eq. (1).

Fig. 4

(a)–(d) Simulation of speckle intensity on the image plane of a single-lens imaging system. The speckle frame size is 90×90  px. Case of (a) no aperture (Video 1, mp4, 9.94 MB [URL: https://doi.org/10.1117/1.JBO.28.4.046005.s1]), a square aperture of (b) 46×46  px (Video 2, mp4, 9.59 MB [URL: https://doi.org/10.1117/1.JBO.28.4.046005.s2]), (c) 12×12  px (Video 3, mp4, 8.26 MB [URL: https://doi.org/10.1117/1.JBO.28.4.046005.s3]), and (d) 4×4  px (Video 4, mp4, 8.54 MB [URL: https://doi.org/10.1117/1.JBO.28.4.046005.s4]). For the sake of simplicity, a square aperture is used instead of a circular aperture. Experimental speckle intensities obtained on (e) a matte surface (Video 5, mp4, 17.5 MB [URL: https://doi.org/10.1117/1.JBO.28.4.046005.s5]) and (f) a Delrin plate (Video 6, mp4, 18.1 MB [URL: https://doi.org/10.1117/1.JBO.28.4.046005.s6]). See Sec. 3.2 for the experimental parameters. In all cases, the applied translational speed is V=10  mm/s.

JBO_28_4_046005_f004.png

2.4.

Correlation Functions

The absolute field correlation function of the stack of simulated speckles was calculated by37

Eq. (18)

g(1)(τ)=E1E2*|E1|2|E2|2,
where E1 and E2 are spatial complex fields at times t1 and t2, respectively. The symbols * and are the complex conjugate and spatial averaging operators, respectively. The time lag corresponding to the frame with a mutual correlation of 1/e with respect to a reference frame was determined as the field correlation time (τc). For the speckle patterns obtained by the experiments, the mutual intensity correlations g(2)(τ) were calculated as

Eq. (19)

g(2)(τ)=I1I2I1I2.
where I1 and I2 are the spatial intensities at times t1 and t2, respectively.

3.

Materials and Methods

3.1.

Scattering Media

In the experimental setup of this study, two static scattering media, namely a matte surface and a Delrin plate of 11-mm thickness, have been used. To prepare the matte surface, a black metal plate was painted 6 times with the intervals of 20 min with Chalk spray (Vintage) of ultra-matte color. The reduced scattering coefficient (μs) of matte is assumed to be infinite.

3.2.

High-Speed Imaging

A continuous wave laser light (CNI MSL-FN-671) of λ=671  nm and a coherence length of longer than 50 m was used to illuminate the scattering media. The laser beam was diverged using a planoconcave lens (Qioptiq, N-BK7, F=6) to realize spherical waves illumination 15 cm away from the scattering medium with a power of 43.8 mW and a beam full-width-at-half-maximum of 15.6 mm.

The single-lens imaging system was made of a hard coated bandpass filter (Edmund Optics) with a wavelength of 675±12.5  nm to minimize background noise, a linear polarizer (Thorlabs, LPNIRE100-B) with the direction perpendicular to polarization of laser light to minimize specular reflection, a circular aperture with the diameter of 2as=0.8  mm, a planoconvex lens (Thorlabs, LA1027-A) of f=3.5  cm, and a high-speed camera (Photron, Fastcam SA3) operated with a frame rate of 10 kHz. The distance from the lens to the camera sensor was set to z2=34.3  mm so that a sharp image was obtained when the matte and Delrin media were located at z1=15.1 and z1=15.4  cm, respectively, from the imaging lens. Here the resulted Fresnel numbers are NF=1.6 and NF=1.5 for matte and Delrin, respectively, indicating that the system is in near-field regime.

The system magnification M (i.e., the fraction of the image size of an object) has been measured as 14. The numerical aperture of the system (NA=as/z2 in air with unity refractive index) is calculated as 0.01. Translational displacement of the scattering medium was realized by mounting them on a linear stage (Zaber, X-LHM200A-E03) perpendicular to the incoming laser beam. Note that the linear stage has a maximum permissible load of 3 kg, which is lower than the mass of the high-speed camera, which is about 6 kg. Therefore, the only option is to move the scattering media mounted on the linear stage.

3.3.

Low-Framerate Imaging

Apart from a high-speed camera, a low-framerate camera meant for a typical LSCI was used to record speckle patterns. For that, an in-house handheld probe was mounted on the Zaber stage in order to apply translational speeds up to 30  mm/s with an acceleration of 3  mm/s2. The matte and Delrin plates were used as the scattering media in this setup. A single-mode optical fiber (Thorlabs SM600, NA=0.12) was used to couple the laser light. The 1/e2   beamwidth at the distance of 20 cm from the fiber tip was measured as 3 cm. A pair of planoconcave and achromatic doublet lenses (f=6  cm) were used to make a collimated beam at the distal end of the optical fiber on the probe. To make scrambled waves, the pair of lenses was replaced by a 20 deg top hat diffuser (Thorlabs EDI-s20-MD).

A monochrome camera (Basler acA2040 55um USB3) at the detection side operated with a framerate of 50 Hz and an exposure time of T=10  ms. The f-number of N=8 for the camera objective (FUJINON HF16XA-5M) with a focal length of f=16  mm was set. We calculate its aperture radius as38

Eq. (20)

as=f2N,
which equals to as=1  mm. Thus the Fresnel number would be NF=7.4 indicating that the system is in the near-field regime. Similar to the high-speed imaging setup, a linear polarizer and a bandpass filter were mounted on the camera objective. Here the magnification was measured as M=0.1. To calculate spatial speckle contrast introduced in Eq. (17), a window of 90×90  px was chosen at the center of scattering spot. This low-framerate LSCI configuration is lighter than 1 kg enabling us to mount it on the linear stage to apply translational motions.

4.

Results

In this section, the proposed model in Sec. 2 has been evaluated and validated using simulations and experiments. The specific parameters of interest are intensity correlation function and contrast of the time integrated speckle intensity. The shape parameters in the speckle simulation are tuned based on the used setups, namely high-speed and low-framerate imaging, so it enables comparing theoretical, experimental, and simulation results. The aperture size is kept constant during this study and since for a one-sided Doppler shift holding paraxial approximation ΔωD=2πVas/λz1, we define {σs,χs}=as/z1. For the high-speed imaging setup, this equals to {σs,χs}=26.5×104 and {σs,χs}=26×104 for matte and Delrin, respectively. For the low-framerate imaging setup, we have χs=5×103 since only the semicircular model of the wavevector density functions are used to simulate speckles.

For the high-speed imaging setup, only spherical waves were used for illumination; therefore, {σi,χi}=0. For the low-framerate imaging setup, planar waves were used for illumination on both matte and Delrin; spherical and scrambled waves were used for illumination of matte only. Thus χi equals zero for both planar and spherical wave illumination schemes. For the scrambled wave illumination on matte, we calculate χi based on the illuminated area on the diffuser by the fiber tip. Assuming d the distance from the fiber tip to the diffuser, χi=dNA/z1. We have performed this experiment for d={13.5,36,56.1}  mm that results in χi={8.1,21.6,33.7}×103, respectively. For the aforementioned combination of illumination types, the optical Doppler histogram was calculated based on Eq. (3).

4.1.

Intensity Correlation Functions

A simulated speckle pattern calculated using Eq. (15) is shown in [Fig. S1(b) in the Supplementary Material and Videos 7 and 8], which is made by a pair of semicircular density functions introduced in Eq. (8) with χi=0.03 and χs=0.01. Note that speckles are simulated directly on the detector and without Fourier filtering as the spatial diffraction (the size of the simulated speckles) does not influence the contrast of the time integrated speckles. Video 7 and Video 8 show the speckle intensity frames during an episode of T=10  ms for V=2  mm/s and V=10  mm/s, respectively, and the corresponding normalized intensity correlation functions g(2)(τ). Here the higher the velocity is, the more dynamic the speckle patterns are, and the more rapid the decay of the intensity correlation will be. In Fig. 5(a), g(2)(τ) versus time lag is shown for a range of applied speeds as well as the theoretical predictions numerically calculated from Eq. (9), which is made by semicircular wavevector density functions for both illumination and detection. The experimental data have been shown in this graph as well, which is carried out using the high-speed imaging setup by spherical illumination on the matte surface.

Fig. 5

Comparison between results of simulated time-varying fully dynamic speckle patterns, experiments and the theory for a high scattering medium (matte). The theoretical curves are obtained by Eqs. (9) and (12) for semicircular and Gaussian wavevector density functions, respectively. Normalized intensity autocorrelations with the (a) semicircular and (b) Gaussian forms of the wavevector density functions. The wavevector density functions for theoretical and simulation data are made by {χi,σi}=0 and {χs,σs}=26.5×104 (Video 7 mp4, 2.37 MB [URL: https://doi.org/10.1117/1.JBO.28.4.046005.s7]; Video 8, mp4, 4.26 MB [URL: https://doi.org/10.1117/1.JBO.28.4.046005.s8]).

JBO_28_4_046005_f005.png

Similarly, dynamic speckles are simulated using Gaussian wavevector density functions, for a range of applied velocities. The calculated intensity correlation functions versus time lag are shown in Fig. 5(b), in which the theoretical g(2)(τ) curves are analytically calculated from Eq. (12). It can be observed that for both semicircular and Gaussian cases, the simulated values are in agreement with the theoretical predictions while there are deviations with the experimental data. Such deviations seem to be less severe in the semicircular case.

4.2.

Speckle Contrast

In a typical LSCI system, the usual measure of the speckle dynamics is spatial contrast introduced in Eq. (17). In Fig. 6(a), the spatial speckle contrast is shown versus the applied translational speed for semicircular density functions introduced in Eq. (8). The theoretical curves are calculated based on Eq. (10), whereas the simulated points are obtained by Eq. (17). Here the detection system (modeled by χs) is kept constant, whereas three different values for the spread of illumination wavevectors (i.e., χi) are realized. χi=0 simulates planar or spherical waves, whereas χi>0 simulates scrambled waves such that a larger χi corresponds to a wider area on a diffuser’s surface being illuminated.30 Similarly, in Fig. 6(b), the spatial speckle contrast is plotted versus the applied translational speed for Gaussian density functions shown in Eq. (11). The theoretical curves are obtained by Eq. (13). The effect of illumination types is simulated here by changing σi while keeping σs unchanged.

Fig. 6

Speckle contrast (simulations) versus the applied translational speed. The density functions for the wavevectors are (a) semicircular and (b) Gaussian. {χi,σi}=0 simulates spherical waves on matte and planar waves on both matte and Delrin, whereas {χi,σi}>0 simulates scrambled waves on matte.

JBO_28_4_046005_f006.png

In Fig. 7, experimental results of the contrast of the time integrated speckles for different illuminations and media types are depicted versus the applied translational speed. Notice that Figs. 7(a)7(c) correspond to the low framerate, whereas Fig. 7(d) corresponds to the high-speed imaging setups.

Fig. 7

Experimental results of the contrast of the time-integrated speckle intensities versus applied translational speed obtained by (a)–(c) low framerate and (d) high-speed imaging setups. Solid curves with the accompanying data points correspond to theoretical and speckle simulation in all subfigures. Data points with error bars represent mean ± standard deviation.

JBO_28_4_046005_f007.png

The case of scrambled wave illumination on a matte surface is shown in Fig. 7(c) for three distance of fiber tip to diffuser d where for a certain speed, the higher the distance, the more the speckle contrast drop is obtained. Here the theoretical curves slightly deviate the corresponding experimental data points for smaller speeds. Finally, in the high-speed imaging setup, an amount of 100 consecutive speckle intensity frames were averaged to realize an exposure time of T=10  ms.

The pair of planar wave illumination on Delrin is depicted in Fig. 7(a), where the experimental data points fairly match the theoretical prediction and simulation. In Fig. 7(b), the two illumination types of planar and spherical waves are shown where the matte surface is used as the scattering medium. The theoretical curves in Figs. 7(a) and 7(b) are the same, and it is expected to obtain the same experimental data for these cases. Then their spatial contrasts were calculated according to Eq. (17) to form the experimental data points shown in Fig. 7(d) versus the applied translational speed for matte and Delrin. For a certain speed, the speckle contrast on Delrin is lower than the speckle contrast calculated on matte. The theoretical curves in Fig. 7 have been numerically calculated using Eq. (10).

5.

Discussion

A theoretical model is proposed to predict MA caused by translation of a handheld LSCI system, for the case of fully static media. The model step by step works back from the speckle contrast to its origins in the optical Doppler shifts, and to the role of the wavevectors of the illumination and the detection that govern these Doppler shifts. Semicircular and Gaussian forms are proposed for the density functions of the incoming and outgoing wavevector sets. Analytical expressions for the contrast of time-integrated intensity for the two types of density functions have been derived that can be calculated numerically.

Equation (7) suggests that the speckle contrast drop is independent of the type of speckle dynamics in terms of boiling, shifting, or a combination of both. This is because in the assumptions of the steps to derive Eq. (7) discussed in Sec. 1 in the Supplementary Material, no discrimination was necessary between boiling or shifting of the speckle movements and the fact that fully developed speckle patterns share equivalent temporal and spatial correlation functions. What dictates the speckle contrast drop is the spatial one-point autocorrelation of the field. This statement is not concluded by the developed model but it is purely originated from the definition of speckle contrast. The importance of this statement is that it facilitates further development of the model.

Prediction of speckle intensity correlation and time-integrated speckle contrast was assessed by simulation of dynamic speckles where an agreement has been found between the simulated and the theoretical values for the cases of semicircular and Gaussian density functions (Figs. 5 and 6). These results are applied for translation of the system when illuminating a high scattering surface with planar, spherical, or scrambled waves as well as translation of the system when illuminating a finite scattering medium with planar waves. For the cases of spherical or scrambled wave illumination on a finite scattering medium, Monte Carlo simulation will be needed to obtain the fluence distribution around each detection point at the surface as an input for the calculation of the optical Doppler distribution A(ωDλ/(2πV)). This forms a crucial topic for the future research.

The experimental study was carried out in two setups, namely, a high-speed and a low-framerate imaging system. In the former, a high scattering surface (matte) and a finite scattering medium (Delrin) were illuminated by spherical waves. Intensity correlation and spatial contrast of the recorded speckle patterns were studied [Figs. 5 and 7(d)]. In the latter, three scenarios were studied: planar wave illumination on Delrin, planar, and spherical wave illuminations on matte and scrambled wave illumination on matte [see Fig. 7(a)7(c)]. It has been observed that planar wave illumination results in the same contrast–speed curve independent of the media [Fig. 7(a) and 7(b)]. This is in agreement with our previous experiments.27 In this study, we have shown that the proposed model can accurately predict, such speckle contrast drop. Also for the same applied speed, a longer diffuser distance to the fiber tip causes more speckle contrast drop due to the role of incoming wavevectors in creation of Doppler shifts [Fig. 7(c)]. Moreover, for the same applied speed, Delrin causes more drop in speckle contrast compared to matte, which is due to the optical wave propagation in the media and the incoming wavevectors being location dependent when using scrambled or spherical waves for illumination [Fig. 7(d)].

The essence of intensity correlation analysis provided in Fig. 5 is first to evaluate theoretical predictions of correlation time derived in Eqs. (9) and (12) by speckle simulation and experiments and second to make a comparison between semicircular and Gaussian forms of the wavevector density functions. As for the theoretical and simulation curves, one might argue that it is the circular reasoning that causes the precise agreement of the corresponding curves because what has been injected as Doppler shift will eventually be extracted in the output. Although both theoretical and simulation cases share the same wavevector density functions, the theoretical curves are obtained directly by calculating Eqs. (9) and (12) and the simulation data are a result of the generation of time-varying speckle intensities and the calculation of their correlation functions based on Eq. (19).

Thus the corresponding pairs would deviate if Eqs. (9) and (12) were derived incorrectly. The same discussion holds for the speckle contrast data shown in Fig. 6, in which the validity of Eqs. (10) and (13) have been examined by calculating contrast of simulated speckles using Eq. (17). As for the comparison between semicircular and Gaussian forms of the wavevector density functions, it is apparent that the semicircular slightly outperforms the Gaussian when it comes to intensity correlation functions (Fig. 5). Therefore, the semicircular form has been used to further calculate the speckle contrast and compare the cases of spherical illumination on matte and Delrin (Fig. 7).

The mismatch between theoretical/simulation and experimental results of intensity correlation obtained in Fig. 5 can be due to experimental imperfections, such as imperfect focus of the lens, aberration, and error in calculation of distances, but also the fact that neither semicircular nor Gaussian are the perfect functions to model the intensity correlation. However, such mismatch becomes negligible when calculating the contrast of time-integrated speckles shown in Fig. 7(d) for the case of Matte. To explain this, we should take into account the weighting factor 1τ/T in Eq. (7). This factor decreases the effect of g(1)(τ) as τ increases especially for the values shorter than 10 ms. Therefore, the simulated and experimental speckle contrast values calculated from this integral will have more agreement.

Although the theoretical/simulated g(2)(τ) start from the value of 2, the experimental curves in Fig. 5 start from lower values. On this basis, the starting points of the theoretical/simulated curves are tuned based on the experimental curves. Such imperfection in the experimental results is due to the specular reflection from the surface, which is minimized using a polarizer but not completely removed. This results in detection of high-intensity values at random locations of a speckle pattern that may not saturate the camera sensor but makes the phase function ϕn(x,y) introduced in Eq. (14) to have a different statistical distribution than uniform. As a result, the so-called partially developed speckle patterns are observed (as opposed to fully developed speckle patterns) and the calculated intensity correlation introduced in Eq. (19) at time lag τ=0 has a value lower than 2. It is also observed in Fig. 5 that the curves of higher speed are of lower g(2)(0). Here the reason is the vibration of the linear stage applied to the pair of illumination-detection system and scattering medium. As the motor rotates faster, more vibrations are made, which causes detection of more dynamic speckle intensities.

We initialized the study of the influence of the spread of ki on the occurring Doppler shifts and the consequent speckle contrast drop using a diffuser for illumination in Ref. 30. In LSCI systems, diffusers are the most common devices used for illumination. In this work, we have studied the influence of ki, ks, and v. We considered single-lens imaging in the experimental setups of this study, which is the most straightforward case to model. Modeling an imaging system enables connecting the experimental parameters, such as aperture size and distances to the scattering surface and image screen to the factors embedded in the model namely the shape parameters. This way, the model is fine-tuned to predict the speckle contrast drop in a system with specific experimental parameters. The other reason to consider the imaging system is to verify that it falls under the Fresnel diffraction by calculating its Fresnel number introduced in (S20) in the Supplementary Material. In this case, the transfer function of the imaging system consists only of the two terms introduced in (S18) and (S19) in the Supplementary Material, which suggests that the speckle correlation function is not influenced by the spatial diffraction. This feature simplifies the procedure of time varying speckle simulation by enabling generation of speckles of size 1 pixel directly on the image screen. Essentially, what control the correlation functions of the speckles are applied speed and the incoming and outgoing wavevector density functions. In Ref. 39, we have employed the model of the current study to predict speckle contrast drop in a handheld LSCI system in which the on-surface speeds have been measured by a vision-based technique in a dual-camera setting.

Throughout this work, only translation of the LSCI system has been studied while in practice, rotation and tilt of wavefront also play role in creation of MA. Also in an in vivo LSCI measurement, the object of interest is human skin, which is dynamic and of finite scattering level. Therefore, it is essential to expand the model to include dynamic media in which some extent of movements including Brownian motion is introduced and that the rotation and tilt of wavefronts are taken into account. In that case, the proposed model should be adjusted and the validation simulation and experiments should be performed.

6.

Conclusion

A model has been developed with the purpose of predicting MA caused by translation in handheld laser speckle contrast perfusion imaging. Our model is based on the optical Doppler shift distributions associated with the range of wavevectors for illumination and detection. It provides a quantitative description of an already introduced influence of illumination, detection, and applied speed on the MA in an LSCI system (Fig. 1).

The model has been validated by simulation of dynamic speckles as well as by experiments with a high-speed camera equipped with a single-lens imaging system as well as a low-framerate camera typical for an LSCI system with a camera objective. The model validation consisted of temporal analysis using intensity correlation functions (Fig. 5) and spatial analysis using contrast of the time-integrated speckle intensity (Figs. 6 and 7). We have concluded that although the model fails to exactly predict the intensity correlation functions, it successfully predicts contrast of the time-integrated speckle when measuring on both a high scattering surface and a tissue mimicking scattering medium (Delrin).

Results of this study help to advance model-based prevention and/or correction of MA occurring during a handheld LSCI measurement. Realization of a handheld LSCI system is crucial in assisting medical staff especially during surgery, in which compactness, speed, and reliability of the measurement system are of pronounced importance.

Disclosures

The authors declare that there are not any financial interests related to this article and no potential conflicts of interest to disclose.

Acknowledgments

This work was supported by the Open Technology Program of the Netherlands Organization for Scientific Research (NWO), Domain Applied and Engineering Sciences (Grant No. 14538). Limited parts of this study have been published in proceedings of SPIE.30,39

Code, Data, and Materials Availability

The dataset generated during this study can be found in dataset 1.40 All the materials used for experiments are in possession of the University of Twente and can be used for the purpose of reproducibility after obtaining permission from the University of Twente.

References

1. 

A. F. Fercher and J. D. Briers, “Flow visualization by means of single-exposure speckle photography,” Opt. Commun., 37 326 –330 https://doi.org/10.1016/0030-4018(81)90428-4 OPCOB8 0030-4018 (1981). Google Scholar

2. 

J. D. Briers, “Laser speckle contrast imaging for measuring blood flow,” Opt. Appl., 37 139 –152 (2007). Google Scholar

3. 

D. A. Boas and A. K. Dunn, “Laser speckle contrast imaging in biomedical optics,” J. Biomed. Opt., 15 011109 https://doi.org/10.1117/1.3285504 JBOPFO 1083-3668 (2010). Google Scholar

4. 

J. Senarathna et al., “Laser speckle contrast imaging: theory, instrumentation and applications,” IEEE Rev. Biomed. Eng., 6 99 –110 https://doi.org/10.1109/RBME.2013.2243140 (2013). Google Scholar

5. 

C. Boudoux, Fundamentals of Biomedical Optics: From Light Interactions with Cells to Complex Imaging Systems, Pollux( (2017). Google Scholar

6. 

S. C. Gnyawali et al., “Retooling laser speckle contrast analysis algorithm to enhance non-invasive high resolution laser speckle functional imaging of cutaneous microcirculation,” Sci. Rep., 7 41048 https://doi.org/10.1038/srep41048 SRCEC3 2045-2322 (2017). Google Scholar

7. 

M. Hultman et al., “Real-time video-rate perfusion imaging using multi-exposure laser speckle contrast imaging and machine learning,” J. Biomed. Opt., 25 116007 https://doi.org/10.1117/1.JBO.25.11.116007 JBOPFO 1083-3668 (2020). Google Scholar

8. 

S. M. S. Kazmi et al., “Flux or speed? Examining speckle contrast imaging of vascular flows,” Biomed. Opt. Express, 6 2588 https://doi.org/10.1364/BOE.6.002588 BOEICL 2156-7085 (2015). Google Scholar

9. 

A. Nadort et al., “Quantitative blood flow velocity imaging using laser speckle flowmetry,” Sci. Rep., 6 25258 https://doi.org/10.1038/srep25258 SRCEC3 2045-2322 (2016). Google Scholar

10. 

D. D. Duncan and S. J. Kirkpatrick, “Can laser speckle flowmetry be made a quantitative tool?,” J. Opt. Soc. Am. A, 25 2088 https://doi.org/10.1364/JOSAA.25.002088 JOAOD6 0740-3232 (2008). Google Scholar

11. 

M. Hultman et al., “Speed-resolved perfusion imaging using multi-exposure laser speckle contrast imaging and machine learning,” J. Biomed. Opt., 28 036007 https://doi.org/10.1117/1.JBO.28.3.036007 JBOPFO 1083-3668 (2023). Google Scholar

12. 

W. Heeman et al., “Clinical applications of laser speckle contrast imaging: a review,” J. Biomed. Opt., 24 080901 https://doi.org/10.1117/1.JBO.24.8.080901 JBOPFO 1083-3668 (2019). Google Scholar

13. 

M. A. Mikhailova et al., “A multimodal approach to monitoring the state of microvasculature in patients with psoriasis in the course of treatment,” Proc. SPIE, 11363 113631N https://doi.org/10.1117/12.2554711 PSISDG 0277-786X (2020). Google Scholar

14. 

M. J. Schaap et al., “Perfusion measured by laser speckle contrast imaging as a predictor for expansion of psoriasis lesions,” Ski. Res. Technol., 28 104 –110 https://doi.org/10.1111/srt.13098 (2022). Google Scholar

15. 

O. A. Mennes et al., “Semi-automatic tracking of laser speckle contrast images of microcirculation in diabetic foot ulcers,” Diagnostics, 10 1054 https://doi.org/10.3390/diagnostics10121054 (2020). Google Scholar

16. 

E. A. Mannoh, “Development of an intraoperative imaging tool for thyroid and parathyroid surgical guidance,” (2021). Google Scholar

17. 

A. Mangraviti et al., “Intraoperative laser speckle contrast imaging for real-time visualization of cerebral blood flow in cerebrovascular surgery: results from pre-clinical studies,” Sci. Rep., 10 7614 https://doi.org/10.1038/s41598-020-64492-5 SRCEC3 2045-2322 (2020). Google Scholar

18. 

R. Farraro, O. Fathi and B. Choi, “Handheld, point-of-care laser speckle imaging,” J. Biomed. Opt., 21 094001 https://doi.org/10.1117/1.JBO.21.9.094001 JBOPFO 1083-3668 (2016). Google Scholar

19. 

D. Amphan, “Quantification and detection of motion artifacts in laser speckle contrast imaging,” (2022). Google Scholar

20. 

B. Lertsakdadet et al., “Handheld motion stabilized laser speckle imaging,” Biomed. Opt. Express, 10 5149 –5158 https://doi.org/10.1364/BOE.10.005149 BOEICL 2156-7085 (2019). Google Scholar

21. 

B. Lertsakdadet et al., “Correcting for motion artifact in handheld laser speckle images,” J. Biomed. Opt., 23 036006 https://doi.org/10.1117/1.JBO.23.3.036006 JBOPFO 1083-3668 (2018). Google Scholar

22. 

L. Omarjee et al., “Optimisation of movement detection and artifact removal during laser speckle contrast imaging,” Microvasc. Res., 97 75 –80 https://doi.org/10.1016/j.mvr.2014.09.005 MIVRA6 0026-2862 (2015). Google Scholar

23. 

G. Mahé et al., “Laser speckle contrast imaging accurately measures blood flow over moving skin surfaces,” Microvasc. Res., 81 183 –188 https://doi.org/10.1016/j.mvr.2010.11.013 MIVRA6 0026-2862 (2011). Google Scholar

24. 

G. Mahé et al., “Cutaneous microvascular functional assessment during exercise: a novel approach using laser speckle contrast imaging,” Pflügers Arch. J. Physiol., 465 451 –458 https://doi.org/10.1007/s00424-012-1215-7 (2013). Google Scholar

25. 

J. Guilbert and M. Desjardins, “Movement correction method for laser speckle contrast imaging of cerebral blood flow in cranial windows in rodents,” J. Biophotonics, 15 e202100218 https://doi.org/10.1002/jbio.202100218 (2022). Google Scholar

26. 

A. Chizari et al., “Exploration of movement artefacts in handheld laser speckle contrast perfusion imaging,” Biomed. Opt. Express, 11 2352 https://doi.org/10.1364/BOE.387252 BOEICL 2156-7085 (2020). Google Scholar

27. 

A. Chizari et al., “Influence of wavefront types on movement artefacts in handheld laser speckle contrast perfusion imaging,” OSA Contin., 4 (6), 1875 –1888 https://doi.org/10.1364/OSAC.420479 (2021). Google Scholar

28. 

A. Chizari et al., “Handheld versus mounted laser speckle contrast perfusion imaging demonstrated in psoriasis lesions,” Sci. Rep., 11 16646 https://doi.org/10.1038/s41598-021-96218-6 SRCEC3 2045-2322 (2021). Google Scholar

29. 

Y. Yeh and H. Z. Cummins, “Localized fluid flow measurements with an HeNe laser spectrometer,” Appl. Phys. Lett., 4 176 –178 https://doi.org/10.1063/1.1753925 APPLAB 0003-6951 (1964). Google Scholar

30. 

A. Chizari et al., “Modeling movement artefacts in handheld laser speckle contrast perfusion imaging: influence of wavefront types,” Proc. SPIE, 11959 1195905 https://doi.org/10.1117/12.2609849 PSISDG 0277-786X (2022). Google Scholar

31. 

I. Fredriksson and M. Larsson, “On the equivalence and differences between laser Doppler flowmetry and laser speckle contrast analysis,” J. Biomed. Opt., 21 126018 https://doi.org/10.1117/1.JBO.21.12.126018 JBOPFO 1083-3668 (2016). Google Scholar

32. 

L. Cohen, “The generalization of the Wiener–Khinchin theorem,” in Proc. 1998 IEEE Int. Conf. Acoust., Speech and Signal Process. (Cat. No. 98CH36181), 1577 –1580 (1998). Google Scholar

33. 

J. G. Proakis, Digital Communications, McGraw-Hill, New York (1989). Google Scholar

34. 

L. Mandel, “Fluctuations of photon beams and their correlations,” Proc. Phys. Soc., 72 1037 https://doi.org/10.1088/0370-1328/72/6/312 (1958). Google Scholar

35. 

J. W. Goodman, Statistical Optics, John Wiley & Sons( (2015). Google Scholar

36. 

F. Bowman, Introduction to Bessel Functions, Dover( (1958). Google Scholar

37. 

D. Ferreira et al., “Connecting field and intensity correlations: the Siegert relation and how to test it,” Am. J. Phys., 88 831 –837 https://doi.org/10.1119/10.0001630 AJPIAS 0002-9505 (2020). Google Scholar

38. 

B. H. Kolner, “Generalization of the concepts of focal length and f-number to space and time,” J. Opt. Soc. Am. A, 11 3229 –3233 https://doi.org/10.1364/JOSAA.11.003229 (1994). Google Scholar

39. 

A. Chizari et al., “Modelling movement artefacts in handheld laser speckle contrast imaging,” Proc. SPIE, 12378 1237804 https://doi.org/10.1117/12.2650617 PSISDG 0277-786X (2023). Google Scholar

40. 

A. Chizari et al., “Dataset for prediction of movement artefacts caused by translation in handheld laser speckle contrast perfusion imaging,” figshare, https://doi.org/10.6084/m9.figshare.21510711 (2023). Google Scholar

Biography

Ata Chizari received his BSc (hons) degree in electronics from Tafresh University, Tafresh, Iran, in 2013 and his MSc (hons) degree in communication systems from Shahid Beheshti University, Tehran, Iran, in 2015. He is a postdoctoral researcher in the Biomedical Photonic Imaging Group of the University of Twente where he received his PhD in 2021 in handheld laser speckle contrast perfusion imaging. From 2014 to 2016, he served as a visitor research assistant at the Optical Networks Research Laboratory, Sharif University of Technology, Tehran, Iran, where he contributed to optical wireless communications and photonics research topics. His research interests include laser speckle contrast imaging, laser Doppler perfusion imaging, and speckle metrology.

Wilson Tsong received his BSc degree in 2019 and his MSc degree in 2022 both in applied physics from the University of Twente. He is currently an engineer at Chilas B.V, a company in Enschede, The Netherlands, which specializes in design and production of lasers. His endeavors are focused in laser optics, in particular research in laser speckle contrast imaging and development of external cavity narrow linewidth tunable lasers.

Tom Knop received his BSc degree in applied physics from Saxion University in Enschede in 2017. For his first internship in the Group Nanobiophysics (NBP) at the University of Twente, he worked on STED nanoscopy. His following and final internship has been in the Biomedical Photonic Imaging Group at the University of Twente, where he helped design and build a two-photon microscope for deep tissue imaging. Currently, he is with BMPI as a technician, working on projects related to microscopy and perfusion imaging.

Wiendelt Steenbergen received his MSc degree in aerospace engineering from Delft University of Technology in 1988, his PhD in fluid dynamics from the Eindhoven University of Technology in 1995, and joined the University of Twente, Ensched, The Netherlands. He is a professor in the Biomedical Photonic Imaging Group of the Faculty of Science and Technology at the University of Twente and a vice-dean research in the same faculty. In 2000, he was appointed as an assistant professor in biomedical optics and broadened his scope to low-coherence interferometry and photoacoustic and acoustooptics imaging for biomedical purposes. In 2010, he became a full professor and a group leader of the Biomedical Photonic Imaging Group of the University of Twente. His current research interests are speckle-based tissue perfusion imaging, combined photoacoustic and ultrasound imaging, tissue perfusion imaging using infrared radiation, and quantification of photoacoustic imaging using acoustooptics.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Ata Chizari, Wilson Tsong, Tom Knop, and Wiendelt Steenbergen "Prediction of motion artifacts caused by translation in handheld laser speckle contrast imaging," Journal of Biomedical Optics 28(4), 046005 (18 April 2023). https://doi.org/10.1117/1.JBO.28.4.046005
Received: 9 November 2022; Accepted: 24 March 2023; Published: 18 April 2023