12 February 2008 Quantitative evaluation of ViOptix's tissue oximeter in an ex-vivo animal model
Jimmy J. M. Mao, Ronald Xu, Bob Lash, Leigh Wright
Author Affiliations +
Abstract
We evaluate the performance of ODISseyTM Tissue Oximeter (ViOptix, Inc., Fremont, CA) against co-oximeter. Concurrent oxygen saturation measurements were made in three dog limbs surgically removed and perfused with an extracorporeal blood circulation system. Oxygen saturation was adjusted in steps ranging from 95% down to 5% as monitored by the co-oximeter. The co-oximeter was used to measure the oxygen saturation of the whole blood drawn from both the arterial and the venous ports of the limb. The tissue oxygenation measured by the ODISseyTM tissue oximeter was compared with the average of the arterial and the venous blood oxygenation measured by the co-oximeter. Linear correlation was observed between the average oxygenation given by the co-oximeter and the ODISseyTM readings, with a root-mean-square difference of 7.6% and the correlation coefficient of 0.941, calculated from N = 194 data points.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jimmy J. M. Mao, Ronald Xu, Bob Lash, and Leigh Wright "Quantitative evaluation of ViOptix's tissue oximeter in an ex-vivo animal model", Proc. SPIE 6849, Design and Quality for Biomedical Technologies, 68490C (12 February 2008); https://doi.org/10.1117/12.751381
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Tissues

Oximeters

Blood

Oxygen

Tissue optics

Near infrared

Back to Top