Paper
7 February 2007 One-camera spectral-domain polarization-sensitive optical coherence tomography
Author Affiliations +
Abstract
We present a simplified spectral-domain polarization-sensitive optical coherence tomography (SD-PS-OCT) setup with a spectrometer that employs only a single line scan camera. Unlike SD-PS-OCT systems reported so far, just one readout of a single line scan camera is necessary to gather information on sample reflectivity, retardation, and optic axis orientation for one A-line. Furthermore, spectrometer alignment and data acquisition synchronization are alleviated by the novel spectrometer design. First in-vivo PS-OCT images of human ocular tissue have been recorded to demonstrate the function of our setup.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bernhard Baumann, Michael Pircher, Erich Götzinger, and Christoph K. Hitzenberger "One-camera spectral-domain polarization-sensitive optical coherence tomography", Proc. SPIE 6429, Coherence Domain Optical Methods and Optical Coherence Tomography in Biomedicine XI, 64290M (7 February 2007); https://doi.org/10.1117/12.702827
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Polarization

Optical coherence tomography

Line scan cameras

Reflectivity

Spectroscopy

In vivo imaging

Tissues